Background: ScanNavTM Anatomy Peripheral Nerve Block (ScanNavTM) is an artificial intelligence (AI)-based device that produces a colour overlay on real-time B-mode ultrasound to highlight key anatomical structures for regional anaesthesia. This study compares consistency of identification of sono-anatomical structures between expert ultrasonographers and ScanNavTM. Methods: Nineteen experts in ultrasound-guided regional anaesthesia (UGRA) annotated 100 structures in 30 ultrasound videos across six anatomical regions. These annotations were compared with each other to produce a quantitative assessment of the level of agreement amongst human experts. The AI colour overlay was then compared with all expert annotations. Differences in human-human and human-AI agreement are presented for each structure class (artery, muscle, nerve, fascia/serosal plane) and structure. Clinical context is provided through subjective assessment data from UGRA experts. Results: For human-human and human-AI annotations, agreement was highest for arteries (mean Dice score 0.88/0.86), then muscles (0.80/0.77), and lowest for nerves (0.48/0.41). Wide discrepancy exists in consistency for different structures, both with human-human and human-AI comparisons; highest for sartorius muscle (0.91/0.92) and lowest for the radial nerve (0.21/0.27). Conclusions: Human experts and the AI system both showed the same pattern of agreement in sono-anatomical structure identification. The clinical significance of the differences presented must be explored; however the perception that human expert opinion is uniform must be challenged. Elements of this assessment framework could be used for other devices to allow consistent evaluations that inform clinical training and practice. Anaesthetists should be actively engaged in the development and adoption of new AI technology.
Ultrasound-guided regional anaesthesia (UGRA) involves the targeted deposition of local anaesthesia to inhibit the function of peripheral nerves. Ultrasound allows the visualisation of nerves and the surrounding structures, to guide needle insertion to a perineural or fascial plane end point for injection. However, it is challenging to develop the necessary skills to acquire and interpret optimal ultrasound images. Sound anatomical knowledge is required and human image analysis is fallible, limited by heuristic behaviours and fatigue, while its subjectivity leads to varied interpretation even amongst experts. Therefore, to maximise the potential benefit of ultrasound guidance, innovation in sono-anatomical identification is required. Artificial intelligence (AI) is rapidly infiltrating many aspects of everyday life. Advances related to medicine have been slower, in part because of the regulatory approval process needing to thoroughly evaluate the risk-benefit ratio of new devices. One area of AI to show significant promise is computer vision (a branch of AI dealing with how computers interpret the visual world), which is particularly relevant to medical image interpretation. AI includes the subfields of machine learning and deep learning, techniques used to interpret or label images. Deep learning systems may hold potential to support ultrasound image interpretation in UGRA but must be trained and validated on data prior to clinical use. Review of the current UGRA literature compares the success and generalisability of deep learning and non-deep learning approaches to image segmentation and explains how computers are able to track structures such as nerves through image frames. We conclude this review with a case study from industry (ScanNav Anatomy Peripheral Nerve Block; Intelligent Ultrasound Limited). This includes a more detailed discussion of the AI approach involved in this system and reviews current evidence of the system performance. The authors discuss how this technology may be best used to assist anaesthetists and what effects this may have on the future of learning and practice of UGRA. Finally, we discuss possible avenues for AI within UGRA and the associated implications.