Although the risk of transfusion-transmitted hepatitis B virus (TT-HBV) infection is very low, it still exists. Therefore, introduction of further precautions to reduce this risk is discussed at present. However, so far no data are available about the HBV vaccination status among blood donors (BDs). We compared HBV vaccination status of apheresis donors (ADs) of a university based and whole BDs (WBDs) of a Red Cross blood donation service using a standardized questionnaire. On the basis of these data, the estimated costs over 10 years for HBV vaccination were calculated for two different strategies and compared with the costs for HBV nucleic acid amplification technology (NAT) testing. 22.3% of the WBDs and 41.2% of the ADs indicated having received at least one HBV vaccine dose. This difference was related to the different demographic structures of the two BD populations (BDPs). With regard to the primary costs for the blood donation service, HBV vaccination of BDs could be an alternative to HBV NAT testing, especially for BDPs with an already high HBV vaccination rate and a high donation frequency.
BACKGROUND: In cancer and transplantation therapy apheresis devices and software of optimum standards are required for the collection of high cell yields with high purity of the desired cell fraction.STUDY DESIGN AND METHODS: In a paired study, 15 healthy blood donors underwent four 10-L leukapheresis procedures (197 33 min) with an inlet blood flow rate of 60 mL per minute by use of two different MNC program settings of the COBE Spectra (Gambro BCT) and the AS.TEC 204 (Fresenius Hemocare) cell separators.RESULTS: Use of the standard MNC program of both apheresis devices resulted in significantly higher (p < 0.01) collection efficiencies of CD14+ monocytes, CD3+ cells, CD4+ cells, CD8+ T cells, CD16+ CD56+ natural killer (NK) cells, and residual PLTs (p < 0.001), owing to higher centrifuge speed. The mean MNC purity of all components was more than 90 percent. By use of standard programs of either device, significant correlations (p < 0.01) between donor monocytes and preleukapheresis NK cell counts and the corresponding component cell yields were found.CONCLUSION: Compared to the program modifications with lower centrifuge velocities the standard MNC programs were significantly more efficient regarding CD14+, CD3+, and CD16+ CD56+ cells. Enhanced centrifuge speed and inlet blood flow rate in MNC programs resulted in higher, similar composed MNC concentrations of the products.