The Low Resolution Imaging Spectrometer (LRIS) for the Cassegrain focus of the Keck 10-meter telescope on Mauna Kea is described. It has an imaging mode so it can also be used for taking direct images. The field of view in both spectrographic and imaging modes is 6 by 7.8 arcmin. It can be used with both conventional slits and custom-punched slit masks. The optical quality of the spectrograph is good enough to take full advantage of the excellent imaging properties of the telescope itself. The detector is a cooled back-illuminated Tektronics Inc. 2048 CCD which gives a sampling rate of 4.685 pixels per arcsec. In the spectrographic mode the spectrograph has a maximum efficiency at the peak of the grating blaze of 32-34% for the two lowest resolution gratings and 28% for the 1200 g/mm grating. This efficiency includes the detector but not the telescope or the atmosphere.
The Low Resolution Imaging Spectrometer is designed for use at the Cassegrain focus of the Keck 10-m telescope. It provides the capability of acquiring low resolution (R equals 1000 to 5000) digital spectra, as well as 6 X 8 arc-minute moderately high spatial resolution (4.65 pixels/arc-second) direct images. Spectroscopy can be carried out with single slits which are 3 arc-minutes long. In addition punched multi-slits can also be employed which allow for the acquisition of at least forty spectra simultaneously. Since the instrument is designed to be as efficient as possible, it is a double spectrograph, with a dichroic splitting the blue and red light into separate optical paths after the collimator. Only the red side has been constructed thus far. With a 2048 by 2048 thinned Tektronix CCD as the detector the total efficiency of the red side at the peak of the grating blaze is predicted to be nearly 40%. Results of the commissioning observing runs will be described.