Despite many potential applications, miniature mass spectrometers have had limited adoption in the field due to the tradeoff between throughput and resolution that limits their performance relative to laboratory instruments. Recently, a solution to this tradeoff has been demonstrated by using spatially coded apertures in magnetic sector mass spectrometers, enabling throughput and signal-to-background improvements of greater than an order of magnitude with no loss of resolution. This paper describes a proof of concept demonstration of a cycloidal coded aperture miniature mass spectrometer (C-CAMMS) demonstrating use of spatially coded apertures in a cycloidal sector mass analyzer for the first time. C-CAMMS also incorporates a miniature carbon nanotube (CNT) field emission electron ionization source and a capacitive transimpedance amplifier (CTIA) ion array detector. Results confirm the cycloidal mass analyzer's compatibility with aperture coding. A >10× increase in throughput was achieved without loss of resolution compared with a single slit instrument. Several areas where additional improvement can be realized are identified. Graphical Abstract ᅟ.
The use of coded apertures in mass spectrometry can break the trade-off between throughput and resolution that has historically plagued conventional instruments. Despite their very early stage of development, coded apertures have been shown to increase throughput by more than one order of magnitude, with no loss in resolution in a simple 90-degree magnetic sector. This enhanced throughput can increase the signal level with respect to the underlying noise, thereby significantly improving sensitivity to low concentrations of analyte. Simultaneous resolution can be maintained, preventing any decrease in selectivity. Both one- and two-dimensional (2D) codes have been demonstrated. A 2D code can provide increased measurement diversity and therefore improved numerical conditioning of the mass spectrum that is reconstructed from the coded signal. This review discusses the state of development, the applications where coding is expected to provide added value, and the various instrument modifications necessary to implement coded apertures in mass spectrometers.
Cycloidal mass analyzers are unique sector mass analyzers as they exhibit perfect double focusing, making them ideal for incorporating spatial aperture coding, which can increase the throughput of a mass analyzer without affecting the resolving power. However, the focusing properties of the cycloidal mass analyzer depend on the uniformity of the electric and magnetic fields. In this paper, finite element simulation and charged particle tracing were used to investigate the effect of field uniformity on imaging performance of a cycloidal mass analyzer. For the magnetic field, we evaluate a new permanent magnet geometry by comparing it to a traditional geometry. Results indicate that creating an aperture image in a cycloidal mass spectrometer with the same FWHM as the slit requires less than 1% variation in magnetic field strength along the ion trajectories. The new magnet design, called the opposed dipole magnet, has less than 1% field variation over an area approximately 62 × 65 mm; nearly twice the area available in a traditional design of similar size and weight. This allows ion imaging across larger detector arrays without loss of resolving power. In addition, we compare the aperture imaging quality of a traditionally used cycloidal mass spectrometer electric design with a new optimized design with improved field uniformity.
The use of microelectromechanical systems (MEMS) components in miniature mass spectrometers is particularly attractive due to their small size and scalable manufacturing capability. Our group has pioneered the development of miniature electron ionization sources combining MEMS fabricated structures with integrated carbon nanotube (CNT) cold-cathode field emitters. However, until now they have been of limited use due to the limited ability to direct the ions into a mass analyzer. In this work, we design a miniature ion source using a microfabricated MEMS device and a low temperature co-fired ceramic (LTCC) carrier that includes electrical connections, ion optics for directing ions out of the device, and a sample inlet. We present the design and fabrication of the ion source; simulate the energy and angular dispersion; and experimentally determine the energy and angular dispersion. (C) 2016 Elsevier B.V. All rights reserved.
L'invention concerne des spectrometres de masse a electrodes segmentees et des procedes associes. Selon un aspect, un appareil ou spectrometre de masse comprend une source d'ions configuree pour generer des ions a partir d'un echantillon. L'appareil comprend egalement un detecteur configure pour detecter une pluralite de rapports masse sur charge associes aux ions. En outre, l'appareil comprend des electrodes segmentees positionnees entre la source d'ions et le detecteur. L'appareil comprend egalement un dispositif de commande configure pour appliquer selectivement une tension aux bornes des electrodes segmentees pour former un profil de champ electrique predetermine.
This work evaluates crosstalk and transmission efficiency in multi-integrated field emission vacuum microelectronic devices (FE-VMDs). Experimental evidence showed that proximity effects cannot be neglected. therefore indicating a need for additional control to take full advantage of the potential microfabricated packing density. Simulations were used to understand the root cause, design structural solutions, and improve overall device performance. Furthermore, charged particle transmission was experimentally investigated for initial structures and new design features are proposed for improved performance.