A method for detection and identification of the hepatitis C virus antigen (HCVcoreAg) in human serum with consideration for possible amino acid substitutions is proposed. The method is based on a combination of biospecific capturing and concentrating of the target protein on the surface of the chip for atomic force microscope (AFM chip) with subsequent protein identification by tandem mass spectrometric (MS/MS) analysis. Biospecific AFM-capturing of viral particles containing HCVcoreAg from serum samples was performed by use of AFM chips with monoclonal antibodies (anti-HCVcore) covalently immobilized on the surface. Biospecific complexes were registered and counted by AFM. Further MS/MS analysis allowed to reliably identify the HCVcoreAg in the complexes formed on the AFM chip surface. Analysis of MS/MS spectra, with the account taken of the possible polymorphisms in the amino acid sequence of the HCVcoreAg, enabled us to increase the number of identified peptides.
A method of atomic force microscopy-based fishing (AFM fishing) has been developed for protein detection in the analyte solution using a chip with an immobilized aptamer. This method is based on the biospecific fishing of a target protein from a bulk solution onto the small AFM chip area with the immobilized aptamer to this protein used as the molecular probe. Such aptamer-based approach allows to increase an AFM image contrast compared to the antibody-based approach. Mass spectrometry analysis used after the biospecific fishing to identify the target protein on the AFM chip has proved complex formation. Use of the AFM chip with the immobilized aptamer avoids interference of the antibody and target protein peaks in a mass spectrum.