BackgroundHigh-quality 3D-anatomy of the day is needed for treatment plan adaptation in radiotherapy. For online x-ray-based CBCT workflows, one approach is to create a synthetic CT or to utilize a fan-beam CT with corresponding registrations. The former potentially introduces uncertainties in the dose calculation if deformable image registration is used. The latter can introduce burden and complexity to the process, the facility, and the patient.PurposeUsing the CBCT of the day, acquired on the treatment device, for direct dose calculation and plan adaptation can overcome these limitations. This study aims to assess the accuracy of the calculated dose on the CBCT scans acquired on a Halcyon linear accelerator equipped with HyperSight.MethodsHyperSight's new CBCT reconstruction algorithm includes improvements in scatter correction, HU calibration of the imager, and beam shape adaptation. Furthermore, HyperSight introduced a new x-ray detector. To show the effect of the implemented improvements, gamma comparisons of 2%/2 mm, 2%/1 mm, and 1%/1 mm were made between the dose distribution in phantoms calculated on the CBCT reconstructions and the simulation CT scans, considering this the standard of care. The resulting gamma passing rates were compared to those obtained with the Halcyon 3.0 reconstruction and hardware without HyperSight's technologies. Various anatomical phantoms for dosimetric evaluations on brain, head and neck, lung, breast, and prostate cases have been used in this study.ResultsThe overall results demonstrated that HyperSight outperformed the Halcyon 3.0 version. Based on the gamma analysis, the calculated dose using HyperSight was closer to the CT scan-based doses than the calculated dose using iCBCT Halcyon 3.0 for most cases. Over all plans and gamma criteria, Halcyon 3.0 achieved an average passing rate of 92.9%, whereas HyperSight achieved 98.1%.ConclusionUsing HyperSight CBCT images for direct dose calculation, for example, in (online) plan adaptation, seems feasible for the investigated cases.
Purpose: Recent evaluations of a 3D iterative cone-beam computed tomography (iCBCT) reconstruction method available on Varian radiation treatment devices demonstrated that iCBCT provides superior image quality when compared to analytical Feldkamp-Davis-Kress (FDK) method. However, iCBCT employs statistical penalized likelihood (PL) that is known to be highly sensitive to inconsistencies due to physiological motion occurring during the acquisition. We propose a computationally inexpensive extension of iCBCT addressing this deficiency. Methods: During the iterative process, the gradients of PL are modified to avoid the generation of motion-related artifacts. To assess the impact of this modification, we propose a motion simulation generating CBCT projections of a moving anatomy together with artifact-free images used as ground truth. Contrast-to-noise ratio and power spectra of difference images are computed to quantify the impact of the motion on reconstructed CBCT volumes as well as the effect of the proposed modification. Results: Using both simulated and clinical data, it is shown that the motion of patient's abdominal wall during the acquisition results in artifacts that can be quantified as low-frequency components in volumes reconstructed with iCBCT. Further, a quantitative evaluation demonstrates that the proposed modification of PL reduces these low-frequency components. While preserving the advantages of PL, it effectively suppresses the propagation of motion-related artifacts into clinically important regions, thus increasing the motion resiliency of iCBCT. Conclusions: The proposed modified iterative reconstruction method significantly improves the quality of CBCT images of anatomies suffering from residual motion.
To evaluate fast-kV switching (FS) dual energy (DE) cone beam computed tomography (CBCT) using the on-board imager (OBI) of a commercial linear accelerator to produce virtual monoenergetic (VM) and relative electron density (RED) images. Using an polynomial attenuation mapping model, CBCT phantom projections obtained at 80 and 140 kVp with FS imaging, were decomposed into equivalent thicknesses of aluminum (Al) and polymethyl methacrylate (PMMA). All projections were obtained with the titanium foil and bowtie filter in place. Basis material projections were then recombined to create VM images by using the linear attenuation coefficients at the specified energy for each material. Similarly, RED images were produced by replacing the linear attenuation values of Al and PMMA by their respective RED values in the projection space. VM and RED images were reconstructed using Feldkamp-Davis-Kress (FDK) and an iterative algorithm (iCBCT, Varian Medical Systems). Hounsfield units (HU), contrast-to-noise ratio (CNR) and RED values were compared against known values. The results after VM-CBCT production showed good material decomposition and consistent HUVM values, with measured root mean square errors (RMSE) from theoretical values, after FDK reconstruction, of 20.5, 5.7, 12.8 and 21.7 HU for 50, 80, 100 and 150 keV, respectively. The largest CNR improvements, when compared to polychromatic images, were observed for the 50 keV VM images. Image noise was reduced up to 28% in the VM-CBCT images after iterative image reconstruction. RED values measured for our method resulted in a mean percentage error of 0.0% ± 1.8%. This study describes a method to generate VM-CBCT and RED images using FS-DE scans obtained using the OBI of a linac, including the effects of the bowtie filter. The creation of VM and RED images increases the dynamic range of CBCT images, and provides additional data that may be used for adaptive radiotherapy, and on table verification for radiotherapy treatments.
We propose an algorithm for periodic motion estimation and compensation in the case of a slowly rotating gantry, e.g., as is the case in cone beam CT. The main target application is abdomen imaging, which is quite challenging because of the absence of high-contrast features. The algorithm is based on minimizing a cost functional, which consists of the data fidelity term, the optical flow constraint term, and regularization terms. To find the appropriate solution we change the constraint strength and regularization strength parameters during the minimization. Results of experiments with simulated and clinical data demonstrate promising performance.
We propose a phase-to-amplitude resampling (PTAR) method to reduce motion blurring in motion-compensated (MoCo) 4D cone-beam CT (CBCT) image reconstruction, without increasing the computational complexity of the motion vector field (MVF) estimation approach. PTAR is able to improve the image quality in reconstructed 4D volumes, including both regular and irregular respiration patterns. The PTAR approach starts with a robust phase-gating procedure for the initial MVF estimation and then switches to a phase-adapted amplitude gating method. The switch implies an MVF-resampling, which makes them amplitude-specific. PTAR ensures that the MVFs, which have been estimated on phase-gated reconstructions, are still valid for all amplitude-gated reconstructions. To validate the method, we use an artificially deformed clinical CT scan with a realistic breathing pattern and several patient data sets acquired with a TrueBeamTM integrated imaging system (Varian Medical Systems, Palo Alto, CA, USA). Motion blurring, which still occurs around the area of the diaphragm or at small vessels above the diaphragm in artifact-specific cyclic motion compensation (acMoCo) images based on phase-gating, is significantly reduced by PTAR. Also, small lung structures appear sharper in the images. This is demonstrated both for simulated and real patient data. A quantification of the sharpness of the diaphragm confirms these findings. PTAR improves the image quality of 4D MoCo reconstructions compared to conventional phase-gated MoCo images, in particular for irregular breathing patterns. Thus, PTAR increases the robustness of MoCo reconstructions for CBCT. Because PTAR does not require any additional steps for the MVF estimation, it is computationally efficient. Our method is not restricted to CBCT but could rather be applied to other image modalities.
PURPOSE Four-dimensional (4D) cone-beam computed tomography (CBCT) of the lung is an effective tool for motion management in radiotherapy but presents a challenge because of slow gantry rotation times. Sorting the individual projections by breathing phase and using an established technique such as Feldkamp-Davis-Kress (FDK) to generate corresponding phase-correlated (PC) three-dimensional (3D) images results in reconstructions (FDK-PC) that often contain severe streaking artifacts due to the sparse angular sampling distributions. These can be reduced by further slowing down the gantry at the expense of incurring unwanted increases in scan times and dose. A computationally efficient alternative is the McKinnon-Bates (MKB) reconstruction algorithm that has shown promise in reducing view aliasing-induced streaking but can produce ghosting artifacts that reduce contrast and impede the determination of motion trajectories. The purpose of this work was to identify and correct shortcomings in the MKB algorithm. METHODS In the general MKB approach, a time-averaged 3D prior image is first reconstructed. The prior is then forward-projected at the same angles as the original projection data creating time-averaged reprojections. These reprojections are subsequently subtracted from the original (unblurred) projections to create motion-encoded difference projections. The difference projections are reconstructed into PC difference images that are added to the well-sampled 3D prior to create the higher quality 4D image. The cause of the ghosting in the traditional 4D MKB images was studied and traced to motion-induced streaking in the prior that, when reprojected, has the undesirable effect of re-encoding for motion in what should be a purely time-averaged reprojection. A new method, designated as the modified McKinnon-Bates (mMKB) algorithm, was developed based on destreaking the prior. This was coupled with a postprocessing 4D bilateral filter for noise suppression and edge preservation (mMKBbf ). The algorithms were tested with the 4D XCAT phantom using four simulated scan times (57, 60, 120, 180 s) and with two in vivo thorax studies (acquisition time of 60 and 90 s). Contrast-to-noise ratios (CNRs) of the target lesions and overall visual quality of the images were assessed. RESULTS Prior destreaking (mMKB algorithm) reduced ghosting artifacts and increased CNRs for all cases, with the biggest impacts seen in the end inhale (EI) and end exhale (EE) phases of the respiratory cycle. For the XCAT phantom, mMKB lesion CNR was 44% higher than the MKB lesion CNR and was 81% higher than the FDK-PC lesion CNR (EI and EE phases). The bilateral filter provided a further average CNR improvement of 87% with the highest increases associated with longer scan times. Across all phases and scan times, the maximum mMKBbf -to-FDK-PC CNR improvement was over 300%. In vivo results agreed with XCAT results. Significantly less ghosting was observed throughout the mMKB images including near the lesions-of-interest and the diaphragm allowing for, in one case, visualization of a small tumor with nearly 30 mm of motion. The maximum FDK-PC-to-MKBbf CNR improvement for Patient 1's lesion was 261% and for Patient 2's lesion was 318%. CONCLUSIONS The 4D mMKB algorithm yields good quality coronal and sagittal images in the thorax that may provide sufficient information for patient verification.
PURPOSE:To correct for scatter in kV cone-beam CT (CBCT) projection data on a clinical system using a new tool, Acuros® CTS, that estimates scatter images rapidly and accurately by deterministically solving the linear Boltzmann transport equation.METHODS:Phantom and patient CBCT scans were acquired on TrueBeam® radiotherapy machines. A first-pass reconstruction was used to create water and bone density maps of the imaged object, which was updated to include a more accurate representation of the patient couch. The imaging system model accounted for the TrueBeam x-ray source (polychromatic spectrum, beam filtration, bowtie filter, and collimation hardware) and x-ray detection system (antiscatter grid, flat-panel imager). Acuros CTS then used the system and object models to estimate the scatter component of each projection image, which was subtracted from the measured projections. The corrected projections were then reconstructed to produce the final result. We examined the tradeoff between run time and accuracy using a Pareto optimization of key parameters, including the voxel size of the down-sampled object model, the number of pixels in the down-sampled detector, and the number of scatter images (angular down-sampling). All computations and reconstructions were performed on a research workstation containing two graphics processing units (GPUs). In addition, we established a method for selecting a subset of projections for which scatter images were calculated. The projections were selected to minimize interpolation errors in the remaining projections. Image quality improvement was assessed by measuring the accuracy of the reconstructed phantom and patient images.RESULTS:The Pareto optimization yielded a set of parameters with an average run time of 26 seconds for scatter correction while maintaining high accuracy of scatter estimation. This was achieved in part by means of optimizing the projection angles that were processed, thus favoring the use of more angles in the lateral (i.e., horizontal) direction and fewer angles in the AP direction. In a 40 cm solid water phantom reconstruction, nonuniformities were decreased from 217 HU without scatter correction to 51 HU with conventional (kernel-based) scatter correction to 17 HU with Acuros CTS-based scatter correction. In clinical pelvis scans, nonuniformities in the bladder were reduced from 85 HU with conventional scatter correction to 14 HU with Acuros CTS.CONCLUSIONS:Acuros CTS is a promising new tool for fast and accurate scatter correction for CBCT imaging. By carefully modeling the imaging chain and optimizing several parameters, we achieved high correction accuracies with computation times compatible with the clinical workflow. The improvement in image quality enables better soft-tissue visualization and potentially enables applications such as adaptive radiotherapy.
To improve the accuracy of motion vector fields (MVFs) required for respiratory motion compensated (MoCo) CT image reconstruction without increasing the computational complexity of the MVF estimation approach, we propose a MVF upsampling method that is able to reduce the motion blurring in reconstructed 4D images. While respiratory gating improves the temporal resolution, it leads to sparse view sampling artifacts. MoCo image reconstruction has the potential to remove all motion artifacts while simultaneously making use of 100% of the rawdata. However the MVF accuracy is still below the temporal resolution of the CBCT data acquisition. Increasing the number of motion bins would increase reconstruction time and amplify sparse view artifacts, but not necessarily the accuracy of MVF. Therefore we propose a new method to upsample estimated MVFs and use those for MoCo. To estimate the MVFs, a modified version of the Demons algorithm is used. Our proposed method is able to interpolate the original MVFs up to a factor that each projection has its own individual MVF. To validate the method we use an artificially deformed clinical CT scan, with a breathing pattern of a real patient, and patient data acquired with a TrueBeamTM4D CBCT system (Varian Medical Systems). We evaluate our method for different numbers of respiratory bins, each again with different upsampling factors. Employing our upsampling method, motion blurring in the reconstructed 4D images, induced by irregular breathing and the limited temporal resolution of phase–correlated images, is substantially reduced.
Kilo-voltage cone-beam computed tomography (CBCT) plays an important role in image guided radiation therapy (IGRT) by providing 3D spatial information of tumor potentially useful for optimizing treatment planning. In current IGRT CBCT system, reconstructed images obtained with analytic algorithms, such as FDK algorithm and its variants, may contain artifacts. In an attempt to compensate for the artifacts, we investigate optimization-based reconstruction algorithms such as the ASD-POCS algorithm for potentially reducing arti- facts in IGRT CBCT images. In this study, using data acquired with a physical phantom and a patient subject, we demonstrate that the ASD-POCS reconstruction can significantly reduce artifacts observed in clinical re- constructions. Moreover, patient images reconstructed by use of the ASD-POCS algorithm indicate a contrast level of soft-tissue improved over that of the clinical reconstruction. We have also performed reconstructions from sparse-view data, and observe that, for current clinical imaging conditions, ASD-POCS reconstructions from data collected at one half of the current clinical projection views appear to show image quality, in terms of spatial and soft-tissue-contrast resolution, higher than that of the corresponding clinical reconstructions.
We propose an adapted method of our previously published five-dimensional (5D) motion compensation (MoCo) algorithm(1), developed for micro-CT imaging of small animals, to provide for the first time motion artifact-free 5D cone-beam CT (CBCT) images from a conventional flat detector-based CBCT scan of clinical patients. Image quality of retrospectively respiratory- and cardiac-gated volumes from flat detector CBCT scans is deteriorated by severe sparse projection artifacts. These artifacts further complicate motion estimation, as it is required for MoCo image reconstruction. For high quality 5D CBCT images at the same x-ray dose and the same number of projections as todays 3D CBCT we developed a double MoCo approach based on motion vector fields (MVFs) for respiratory and cardiac motion. in a first step our already published four-dimensional (41)) artifact-specific cyclic motion-compensation (acMoCo) approach is applied to compensate for the respiratory patient motion. With this information a cyclic phase-gated deformable heart registration algorithm is applied to the respiratory motion-compensated 4D CBCT data, thus resulting in cardiac MVFs. We apply these MVFs on double-gated images and thereby respiratory and cardiac motion-compensated 5D CBCT images are obtained. Our 5D MoCo approach processing patient data acquired with the TrueBeam 4D CBCT system (Varian Medical Systems). Our double MoCo approach turned out to be very efficient and removed nearly all streak artifacts due to making use of 100% of the projection data for each reconstructed frame. The 5D MoCo patient data show fine details and no motion blurring, even in regions close to the heart where motion is fastest.
PURPOSE Implanted gold markers for image-guided radiotherapy lead to streaking artifacts in cone-beam CT (CBCT) scans. Several methods for metal artifact reduction (MAR) have been published, but they all fail in scans with large motion. Here the authors propose and investigate a method for automatic moving metal artifact reduction (MMAR) in CBCT scans with cylindrical gold markers. METHODS The MMAR CBCT reconstruction method has six steps. (1) Automatic segmentation of the cylindrical markers in the CBCT projections. (2) Removal of each marker in the projections by replacing the pixels within a masked area with interpolated values. (3) Reconstruction of a marker-free CBCT volume from the manipulated CBCT projections. (4) Reconstruction of a standard CBCT volume with metal artifacts from the original CBCT projections. (5) Estimation of the three-dimensional (3D) trajectory during CBCT acquisition for each marker based on the segmentation in Step 1, and identification of the smallest ellipsoidal volume that encompasses 95% of the visited 3D positions. (6) Generation of the final MMAR CBCT reconstruction from the marker-free CBCT volume of Step 3 by replacing the voxels in the 95% ellipsoid with the corresponding voxels of the standard CBCT volume of Step 4. The MMAR reconstruction was performed retrospectively using a half-fan CBCT scan for 29 consecutive stereotactic body radiation therapy patients with 2-3 gold markers implanted in the liver. The metal artifacts of the MMAR reconstructions were scored and compared with a standard MAR reconstruction by counting the streaks and by calculating the standard deviation of the Hounsfield units in a region around each marker. RESULTS The markers were found with the same autosegmentation settings in 27 CBCT scans, while two scans needed slightly changed settings to find all markers automatically in Step 1 of the MMAR method. MMAR resulted in 15 scans with no streaking artifacts, 11 scans with 1-4 streaks, and 3 scans with severe streaking artifacts. The corresponding numbers for MAR were 8 (no streaks), 1 (1-4 streaks), and 20 (severe streaking artifacts). The MMAR method was superior to MAR in scans with more than 8 mm 3D marker motion and comparable to MAR for scans with less than 8 mm motion. In addition, the MMAR method was tested on a 4D CBCT reconstruction for which it worked equally well as for the 3D case. The markers in the 4D case had very low motion blur. CONCLUSIONS An automatic method for MMAR in CBCT scans was proposed and shown to effectively remove almost all streaking artifacts in a large set of clinical CBCT scans with implanted gold markers in the liver. Residual streaking artifacts observed in three CBCT scans may be removed with better marker segmentation.
A generic model-based segmentation algorithm is presented. Based on a set of training data, consisting of images with corresponding object segmentations, a local appearance and local shape model is build. The object is described by a set of landmarks. For each landmark a local appearance model is build. This model describes the local intensity values in the image around each landmark. The local shape model is constructed by considering the landmarks to be vertices in an undirected graph. The edges represent the relations between neighboring landmarks. By implying the markovianity property on the graph, every landmark is only directly dependent upon its neighboring landmarks, leading to a local shape model. The objective function to be minimized is obtained from a maximum a-posteriori approach. To minimize this objective function, the problem is discretized by considering a finite set of possible candidates for each landmark. In this way the segmentation problem is turned into a labeling problem. Mean field annealing is used to optimize this labeling problem. The, algorithm is validated for the segmentation of teeth from cone beam computed tomography images and for automated cephalometric analysis.
This paper presents a comparison study between 10 automatic and six interactive methods for liver segmentation from contrast-enhanced CT images. It is based on results from the "MICCAI 2007 Grand Challenge" workshop, where 16 teams evaluated their algorithms on a common database. A collection of 20 clinical images with reference segmentations was provided to train and tune algorithms in advance. Participants were also allowed to use additional proprietary training data for that purpose. All teams then had to apply their methods to 10 test datasets and submit the obtained results. Employed algorithms include statistical shape models, atlas registration, level-sets, graph-cuts and rule-based systems. All results were compared to reference segmentations five error measures that highlight different aspects of segmentation accuracy. All measures were combined according to a specific scoring system relating the obtained values to human expert variability. In general, interactive methods reached higher average scores than automatic approaches and featured a better consistency of segmentation quality. However, the best automatic methods (mainly based on statistical shape models with some additional free deformation) could compete well on the majority of test images. The study provides an insight in performance of different segmentation approaches under real-world conditions and highlights achievements and limitations of current image analysis techniques.
A generic supervised segmentation approach is presented. The object is described as a graph where the vertices correspond to landmarks points and the edges define the landmark relations. Instead of building one single global shape model, a priori shape information is represented as a concatenation of local shape models that consider only local dependencies between connected landmarks. The objective function is obtained from a maximum a posteriori criterion and is build up of localized energies of both shape and landmark intensity information. The optimization problem is discretized by searching candidates for each landmark using individual landmark intensity descriptors. The discrete optimization problem is then solved using mean field annealing or dynamic programming techniques. The algorithm is validated for hand bone segmentation from RX datasets and for 3D liver segmentation from contrast enhanced CT images.
Liver segmentation is an important step for the therapeutic decision making in liver surgery. However, manual segmentation is timeconsuming and tedious and so the need for accurate and robust automatic segmentation methods for clinical data arises. In this work an atlas in combination with nonrigid registration is used to segment the liver in actual clinical CT images. First, the atlas is built on twenty training images using nonrigid registration with a novel surface distance penalty. Next, this atlas is nonrigidly registered to ten test images. Currently, the user interaction is limited to the initialization of a rigid registration and to the definition of a region of interest for the nonrigid registration. Future work will focus on replacing the remaining user interaction with fully automatic procedures. Results are promising with an average overlap error of 10.4% and an average RMS distance of 5.0mm for the ten test images. Errors occur mainly at sites where the atlas is ill-defined such as the border between the heart and the liver.
A 3D medical image segmentation algorithm is presented which models an object as a set of landmarks augmented with local appearance models similar as with Active Shape Model segmentation methods, but instead of applying a global shape model, multiple local shape models are used. A global cost function incorporating local in- tensity and local shape is optimized iteratively. A first implementation of the method is validated for the segmentation of the liver in contrast enhanced CT images and demonstrates that the method has potential. This work participates in the MICCAI 2007 Grand Challenge on 3D segmentation (1).