A method for the extraction of clenbuterol from calf urine samples using a molecularly imprinted polymer (MIP) has been developed. The aim was that the final extracts from the MIP should allow quantitation of clenbuterol down to 0.5 ng/mL urine using HPLC with UV detection. The MIP was produced using brombuterol as a template and the selectivity of the MIP, for clenbuterol, was tested against a non-imprinted polymer (produced without template) and was found to be high. After loading of 5 mL diluted centrifuged urine, selective binding was established in acetonitrile-acetic acid (98:2). For further elution of interferences, 0.5 M ammonium acetate buffer pH 5 and 70% acetonitrile in water was used. Clenbuterol was eluted using 1% trifluoroacetic acid in methanol, which was evaporated and reconstituted in buffer. Results from the HPLC analyses showed that the extraction of clenbuterol using MIP is linear in the range 0.5-100 ng/mL with good precision (4.3% for 0.6 ng/mL and 2.1% for 6.0 ng/mL) and accuracy (96.7% for 0.6 ng/mL and 96.7% for 6.0 ng/mL). The recoveries were 75%. The results show that the method offers a selectivity and sensitivity that make the quantitation of 0.5 ng clenbuterol/mL urine by HPLC-UV possible and a competitive alternative to state-of-the-art routine analytical methods.
An interlaboratory pilot study was performed to determine the reproducibility of mobility parameters in capillary zone electrophoresis (CZE) and micellar electrokinetic chromatography (MEKC). The study was performed by an intended small number of laboratories (three) that used different brands of instruments (two). The effective mobility was corrected using standards by a method that was recently introduced to obtain a more reproducible migration parameter. A test set of 20 acidic test compounds and 5 reference compounds were analyzed during five days in each laboratory using CZE and MEKC. Buffers used consisted of 90 mM borate set at pH 8.4 (CZE) and 20 mM phosphate, 50 mM sodium dodecyl sulfate set at pH 7.5 (MEKC). Analyses were carried out using fused-silica capillaries at an electric field strength of either 52.6 kV/m or 37.5 kV/m. The interlaboratory reproducibility (mean RSD) of the effective mobility was 3.0% for CZE and 6.7% for MEKC. After applying the correction method, these values became 3.0% for CZE and 3.3% for MEKC, which is adequate for systematic toxicological analysis (STA) applications. A significant improvement of reproducibility for the calculated corrected effective mobility mu(eff)c was observed when variations are high. Therefore, it is recommended to use the correction method in interlaboratory situations, especially when instruments and capillaries from different manufacturers are used.
Spherical molecularly imprinted polymer particles obtained via precipitation polymerization, were introduced as a pseudostationary phase in capillary electrophoresis (CE) to study molecular recognition. Analyses were performed via a partial filling technique using (+)-ephedrine-imprinted microspheres (100-200 nm) which were polymerized from methacrylic acid and 1,1,1-Tris(hydroxymethyl)propanetrimethacrylate using acetonitrile as the solvent. The influence of pH and the modifier content on the separation was investigated. A 0.1% w/v suspension in an aqueous 10 mM phosphate buffer (pH 2.5 with 40% acetonitrile) was hydrodynamically injected into the CE system (80% of the effective capillary length) and led to full baseline separation of racemic ephedrine within 10 min.
Highly selective sorbents for solid-phase extraction are desirable tools for the development of selective and sensitive methods for trace analysis. Molecular imprinting is an emerging technique that can yield analyte- or group-selective polymeric sorbents. Guest authors Ensing and Berggren and column editor Majors discuss the basic principles of molecular imprinting and compare the technique with conventional sorbent and immunoaffinity-based solid-phase extraction. They use real-life examples to illustrate the potential of molecularly imprinted polymer-based SPE.
This article describes a novel nonisotopic receptor assay for benzodiazepines with fluorescence detection. As labeled ligand (coumarin-labeled desethylflumazenil, CLDEF), a metabolite of the benzodiazepine antagonist flumazenil (desetheylflumazenil, Ro15-3890) has been coupled to a coumarin fluorophore, via a spacer. CLDEF had a Ki of 6.5 nM. To avoid the interference of the background fluorescence of the receptors in the measurement step, the bound CLDEF was dissociated from the receptors after the filtration step. This dissociation was achieved by incubating the CLDEF-bound to the receptors on the filters-with a weakly acetate buffer. The second filtrates then contained the previously bound CLDEF, which was then quantitated with a RP-HPLC system with a fluorescence detector. The results with a fluorescent receptor assay were very similar to those with a radioreceptor assay, in that the IC50 values of lorazepam were 7.2 +/- 0.5 and 6.6 +/- 0.7 nM, respectively.
Matrix solid-phase dispersion (MSPD) is a new sample pretreatment for solid samples. This technique greatly simplifies sample pretreatment but, nonetheless, the extracts often still require an extra cleanup step that is both laborious and time-consuming. The potential of combining MSPD with molecularly imprinted solid-phase extraction (MISPE) was investigated in this study. Liver samples were ground in a mortar with C18 sorbent and the homogenized mixture packed into an SPE cartridge and placed on top of a MISPE cartridge. Subsequently, clenbuterol was eluted from the MSPD cartridge onto the MISPE cartridge using acetonitrile containing 1% acetic acid. The ability of the molecularly imprinted polymer to selectively adsorb analyte in acetonitrile was exploited for re-extracting clenbuterol directly from this acetonitrile extract via the double cartridge tandem system. The analyte was eluted from the MISPE cartridge using acidified methanol. A clear eluate was obtained, which was subsequently evaporated, redissolved, and analyzed by HPLC electrochemical detection (ECD) or ion trap mass spectrometry (LC/IT-MS). The MISPE cartridge used in this study was imprinted using bromoclenbuterol, a structural analogue of clenbuterol, as the template. These MISPE cartridges showed excellent stability. The complete extraction procedure was rapid, and recoveries exceeded 90% for the target analyte. The method detection limit for the LC/IT-MS procedure was < 0.1 microg/kg. This method, therefore, satisfies the stringent requirements of European Union regulation EEC 2377/90.
Polymers imprinted with clenbuterol were used to study the influence of various post-polymerization treatments [e.g., thermal annealing, microwave assisted extraction (MAE), Soxhlet extraction and supercritical fluid template desorption] on the bleeding of residual template. The aim of the study was to reduce the bleeding to levels that would allow the use of the materials as affinity phases for extraction of clenbuterol from bovine urine at concentrations below 1 ng ml-1. After treatment, the clenbuterol imprinted polymers were packed into solid-phase extraction columns and the bleeding was estimated by quantifying the amount of template released in 10 ml of methanol-acetic acid (9 + 1 v/v). This was followed by an assessment of selectivity and recovery in comparison with non-treated material. The lowest bleeding level was found after MAE using 100% trifluoroacetic acid for 3 x 20 min at 100 degrees C. The collected eluate contained in this case 3 ng ml-1 of clenbuterol. The same material was subsequently used for the extraction of clenbuterol from spiked bovine urine. The resulting selectivity and recovery were lower compared with those obtained using the untreated material. A milder but still efficient method to reduce the bleeding level was found to be MAE with formic acid. In this case a bleeding level of 14 ng ml-1 was found after only a 1 h extraction time. In a second model system, using a polymer imprinted with L-phenylalanine anilide, the bleeding was reduced to a similar level by extensive on-line washing in good swelling solvents containing acid or base additives and after thermal annealing of the polymers in the dry state.
Capillary electrophoresis (CE) is a modern separation technique that has some distinct advantages for toxicological analysis, such as a high efficiency, fast analysis, flexibility, and complementary separation mechanisms to chromatographic methods. CE can be applied in various modes, which each have a different separation mechanism or selectivity. The most common mode is capillary zone electrophoresis (CZE), in which charged analytes migrate in a buffer under the influence of an electric field. In micellar electrokinetic chromatography (MEKC), micelles are added to the buffer which interact with the analytes. MEKC can also be used for the separation of neutral compounds. In non-aqueous CE (NACE), the aqueous buffer is replaced by a background of electrolytes in organic solvents. A sample that needs to be screened can easily be analyzed subsequently by these CE modes using the same instrumentation. The aim of the study was to develop procedures for the analysis of basic and acidic drugs in serum and urine using CZE, MEKC, and NACE. A test mixture that consisted of six basic and six acidic compounds was used to study the separation behavior of five CE methods. The results showed that three methods (based on CZE, MEKC, and NACE) were suitable for the analysis of basic compounds and three methods (based on CZE and MEKC) for the analysis of acidic compounds. For the extraction of analytes from serum and urine, a solid-phase extraction (SPE) and a liquid-liquid extraction (LLE) method were compared. Both SPE and LLE methods provided clean extracts after extraction of the basic compounds from serum and urine. The extracts of acidic compounds contained more matrix interferences, especially for urine. The SPE method had some advantages compared to LLE, as it lead to cleaner extracts and higher peaks, and as it elutes basic and acidic compounds in one fraction. The potentials and pitfalls of the various methods for screening purposes in analytical toxicology are discussed.
A method was developed for the enantioseparation of ofloxacin, a member of the fluoroquinolones, using an anionic cyclodextrin-derivative with or without combination with a neutral cyclodextrin-derivative, as the chiral selector (s) in an electrokinetic chromatography system. The best results were obtained with 0.35 mM sulfated beta-cyclodextrin dissolved in a 50 mM phosphate buffer, pH 2.5, and at 15 degrees C. Under these conditions, a resolution of 2 was readily achieved. Furthermore, under adequate separation conditions, studies were performed in order to assess possible in vitro and in vivo enantioconversion of levofloxacin. The current method allows detection of 2 microg R-(+)-ofloxacine/mL diluted urine without the necessity of sample cleanup.
The simplicity and flexibility of solid-phase microextraction have been combined with the selectivity of molecularly imprinted polymers (MIPs). Silica fibers were coated reproducible with a 75-microm layer of methacrylate polymer either nonimprinted or imprinted with clenbuterol to compare their extraction characteristics under various conditions. Although the template molecule could be removed effectively from the imprinted polymer, structural analogues of clenbuterol were used for evaluation. The influence of pH on the extractability of brombuterol was investigated. Extraction yields up to approximately 80% were obtained when both types of fibers were used to extract brombuterol from phosphate buffer (pH 7.0). In contrast, yields of about 75 and <5% were obtained when extraction was performed from acetonitrile with imprinted and nonimprinted polymers, respectively, which demonstrates the selectivity of the MIP-coated fiber. Time sorption profiles were measured for the extraction of brombuterol from buffer and acetonitrile at the 10 and 100 ng/mL level with both types of fibers in order to compare extraction characteristics. Equilibrium times of about 30 and 90 min were found for the extraction of brombuterol from acetonitrile and buffer, respectively. The MIP-coated fibers were capable of extracting five structural analogues of clenbuterol from both buffer and acetonitrile, which suggests that the amine alcohol part of these molecules is responsible for interaction with the imprinted polymer. To achieve selective extraction of brombuterol from human urine, MIP-coated fibers were washed with acetonitrile after the extraction. Clean extracts and yields of approximately 45% were obtained, demonstrating the suitability of MIP-coated fibers for the analysis of biological samples.
In systematic toxicological analysis (STA), analytical methods should have a high identification power. This can be suitably expressed by parameters such as mean list length (MLL) or discriminating power (DP). The reproducibility of a method has a great impact on its identification power, and should be as high as possible. In this study, two separation methods based on capillary zone electrophoresis (CZE) were evaluated towards STA applications. Besides a normal phosphate buffer, the commercially available buffer CElixir was used, which is a double-layer dynamic coating system. The coating stabilizes the endoosmotic flow, is independent of the pH, and is claimed to be more reproducible and faster at low pH than with normal buffers. A test set of 73 basic pharmaceutical compounds was analyzed by the two CZE methods. The total analysis time, including rinsing steps, was 8 min when the coating was used and 18 min without the coating. Effective mobilities were calculated and the reproducibilities were a factor of 2 better when the coating was used (between-days SD 0.020 and 0.040 m2/V s with and without the coating, respectively). MLL and DP were calculated for the two CZE methods and for combinations with standardized liquid and gas chromatography systems. CZE with CElixir coating clearly has a high potential for STA applications, as it was shown to have a higher identification power and shorter analysis times than normal CZE.
In this article, the binding affinities of the fluorescent-labeled benzodiazepines described in Part 1 are compared to assess the influence of the labeling position and the choice of fluorophore on the binding affinity. This comparison was extended by taking into account the data of other fluorescent-labeled benzodiazepines in the literature. The differences in the binding affinities observed could partly be explained by structure-activity relationships (SAR). On the basis of this comparison, fluorescent-labeled desethylflumazenil (Ro15-3890, 19) derivatives were selected as the most suitable labeled ligands in fluorescent receptor assays. A methyl-methoxycoumarin derivative (Mmc-O-CO-(CH2)3-Ro15-3890) (20b) had a Ki-value of 6.5 nM, and a 7-nitrobenz-2-oxa-1,3diazole derivative (NBD-NH-(CH2)3-Ro15-3890, 21) had a Ki-value of 5.7 nM. In order to yield sufficient sensitivity in the final receptor assay, a suitable fluorescent labeled ligand should have a Ki < 10 nM. A further advantage of the above two ligands is that the benzodiazepine moiety has no receptor affinity of its own. Thus, if some hydrolysis of the labeled ligand were to occur, the resulting Ro15-3890 (18) would hardly affect the outcome of the assay. In the second part of this paper the prerequisites of the fluorophore are being examined. In this regard, 20b is preferred, because the coumarin derivative has higher fluorescence intensities in aqueous media than the NBD-derivative. Therefore, 20b was selected as a fluorescent-labeled ligand in the development of a non-radioactive receptor assay for benzodiazepines.
A review is presented on the use of charged cyclodextrins (CDs) as chiral selectors in capillary electrophoresis (CE) for the separation of analytes in pharmaceutical analysis. An overview is given of theoretical models that have been developed for a better prediction of the enantiomeric resolution and for a better understanding of the separation mechanism. Several types of charged CDs have been used in chiral capillary electrophoretic separation (anionic, cationic, and amphoteric CDs). Especially the anionic CDs seem to be valuable due to the fact that many pharmaceutically interesting compounds can easily be protonated (e.g., amine groups). For that reason several anionic CDs are now commercially available. Cationic and amphoteric CDs are less common in chiral analysis and only a few are commercially available. Attention is paid to the most common synthesis routes and the characterization of the CDs used in chiral capillary electrophoretic separations. The degree of substitution in the synthesized CDs may vary from one manufacturer to another or even from batch to batch, which may have a detrimental effect on the reproducibility and ruggedness of the separation system. In Sections 4, 5, and 6 the applications of anionic, cationic, and amphoteric CDs for the chiral separation in CE are described. Many interesting examples are shown and the influence of important parameters on the enantioselectivity is discussed.
A feasibility study was performed in order to study the possibilities in using molecularly imprinted polymers (MIPs) as sorbent material in solid-phase extraction (MISPE) for clean-up of clenbuterol from urine. A binding study of clenbuterol in several solvents was performed on a clenbuterol imprinted polymer as well as on a blank polymer. These binding experiments were used to find suitable loading, washing and elution solvents for the MISPE procedure. Extraction of clenbuterol from calf urine was performed by directly loading a 10-ml urine sample onto the MIP column. Thereafter the column was washed with 10 ml of acetonitrile containing 1% acetic acid, and finally clenbuterol was eluted with 6 ml of methanol containing 10% acetic acid. A recovery of 65% was obtained. This recovery could be increased up to 75% if a sample volume of 1 ml was used or up to 100% if urine was freeze-dried and the residue was dissolved in acetonitrile and spiked with clenbuterol prior to analysis. Chromatograms of the wash and eluate solutions show an efficient clean-up, which supports the potential of MISPE for clean-up of trace amounts of clenbuterol from calf urine.
Because radioactive labeled ligands in receptor assays have several disadvantages, we synthesized a number of fluorescent-labeled benzodiazepines. Several fluorophores were attached at different positions of 1,4-benzodiazepine molecules in order to assess the impact of the fluorophores and their coupling position on the affinity for the benzodiazepine receptor. Besides the 1,4-benzodiazepines, the 1,2-annelated 1,4-benzodiazepines were also used for labeling. A metabolite of flumazenil (18), desethylflumazenil (Ro15-3890, 19), was labeled with the fluorophore 4-bromomethyl-7-methoxycoumarin, with and without the incorporation of a spacer chain, yielding the methyl-methoxycoumarin (Mmc) derivatives Mmc-Ro15-3890 (20a) and Mmc-O-CO-(CH2)3-Ro15-3890 (20b), respectively. After the synthesis, the fluorescent-labeled benzodiazepines were purified by HPLC, using an analytical RP-C18 column. For the purification of 20b, the chromatographic system was optimized, using multi-criteria decision making (MCDM) techniques. The binding affinities for the benzodiazepine receptor and the fluorescence characteristics were determined for the resulting products.
The intra- and interinstrument reproducibilities of four capillary electrophoresis instruments were studied for identification purposes in systematic toxicological analysis (STA). A test set of 20 acidic test compounds and 5 reference compounds were analyzed for five days on each instrument using capillary zone electrophoresis (CZE) and micellar electrokinetic chromatography (MEKC). The buffers consisted of 90 mM borate set at pH 8.4 (CZE) and 20 mM phosphate and 50 mM sodium dodecyl sulfate set at pH 7.5 (MEKC). All analyses were carried out using fused silica capillaries at an electric field strength of 52.6 kV/m. The use of a reproducible identification parameter is very important in STA. To deal with the poor reproducibility of the migration time, we recently introduced the corrected effective mobility. In this study, we investigated the intra- and interinstrument reproducibility of the migration time, the effective mobility, and the corrected effective mobility. Large differences in intra-instrument reproducibility were found when the migration time was used. The calculation of the effective mobility and the corrected effective mobility diminished these differences and enhanced the interinstrument reproducibility roughly by a factor 3. For (corrected) effective mobilities, intrainstrument reproducibilities were between 0.8-2.6% and interinstrument reproducibilities were between 3.2-3.9%.