Purpose:The imaging of pregnant patients is medically necessary in certain clinical situations. The purpose of this work was to directly measure uterine doses in a cadaver scanned with CT protocols commonly performed on pregnant patients in order to estimate fetal dose and assess potential risk.Method:One postmortem subject was scanned on a 320‐slice CT scanner with standard pulmonary embolism, trauma, and appendicitis protocols. All protocols were performed with the scan parameters and ranges currently used in clinical practice. Exams were performed both with and without iterative reconstruction to highlight the dose savings potential. Optically stimulated luminescent dosimeters (OSLDs) were inserted into the uterus in order to approximate fetal doses.Results:In the pulmonary embolism CT protocol, the uterus is outside of the primary beam, and the dose to the uterus was under 1 mGy. In the trauma and appendicitis protocols, the uterus is in the primary beam, the fetal dose estimates were 30.5 mGy for the trauma protocol, and 20.6 mGy for the appendicitis protocol. Iterative reconstruction reduced fetal doses by 30%, with uterine doses at 21.3 for the trauma and 14.3 mGy for the appendicitis protocol.Conclusion:Fetal doses were under 1 mGy when exposed to scatter radiation, and under 50 mGy when exposed to primary radiation with the trauma and appendicitis protocols. Consistent with the National Council on Radiation Protection & Measurements (NCRP) and the International Commission on Radiological Protection (ICRP), these doses exhibit a negligible risk to the fetus, with only a small increased risk of cancer. Still, CT scans are not recommended during pregnancy unless the benefits of the exam clearly outweigh the potential risk. Furthermore, when possible, pregnant patients should be examined on CT scanners equipped with iterative reconstruction in order to keep patient doses as low as reasonable achievable.
PURPOSE To characterize the distribution of the deviation index (DI) in digital radiography practices across the United States. METHODS DI data was obtained from 10 collaborating institutions in the United States between 2012 and 2015. Each institution complied with the requirements of the Institutional Review Board at their site. DI data from radiographs of the body parts chest, abdomen, pelvis and extremity were analyzed for anteroposterior, posteroanterior, lateral, and decubitus views. The DI data was analyzed both in aggregate and stratified by exposure control method, image receptor technology, patient age, and participating site for each body part and view. The number of exposures with DI falling within previously published control limits for DI and descriptive statistics were calculated. RESULTS DI data from 505,930 radiographic exposures was analyzed. The number of exposures with DI falling within published control limits for DI varied from 10 to 20% for adult patients and 10 to 23% for pediatric patients for different body parts and views. Mean DI values averaged over other parameters for radiographs of the abdomen, chest, pelvis, and extremities ranged from 0.3 to 1.0, -0.6 to 0.5, 0.8, and -0.9 to 0.5 for the different adult views and ranged from -1.6 to -0.1, -0.3 to 0.5, -0.1, -0.2 to 1.4 for the different pediatric views, respectively (DI data was solicited only for anteroposterior view of pelvis). Standard deviation values of DI from individual sites ranged from 1.3 to 3.6 and 1.3 to 3.0 for the different adult and pediatric views, respectively. Also of interest was that target exposure indicators varied by up to a factor of 6 between sites for certain body parts and views. CONCLUSION Previously published DI control limits do not reflect the state of clinical practice in digital radiography. Mean DI and target exposure indicators are targets for quality improvement efforts in radiography.
Purpose:To optimize adult head CT protocol by reducing dose to an appropriate level while providing CT images of diagnostic quality.Methods:Five cadavers were scanned from the skull base to the vertex using a routine adult head CT protocol (120 kVp, 270 mA, 0.75 s rotation, 0.5 mm × 32 detectors, 70.8 mGy CTDIvol) followed by seven reduced‐dose protocols with varying combinations of reduced tube current, reduced rotation time, and increased detectors with CTDIvol ranging from 38.2 to 65.6 mGy. Organ doses were directly measured with 21 OSL dosimeters placed on the surface and implanted in the head by a neurosurgeon. Two neuroradiologists assessed grey‐white matter differentiation, fluid space, ventricular size, midline shift, brain mass, edema, ischemia, and skull fractures on a three point scale: (1) Unacceptable, (2) Borderline Acceptable, and (3) Acceptable.Results:For the standard scan, doses to the skin, lens of the eye, salivary glands, thyroid, and brain were 37.55 mGy, 49.65 mGy, 40.67 mGy, 4.63 mGy, and 27.33 mGy, respectively. Two cadavers had cerebral edema due to changing dynamics of postmortem effects, causing the grey‐white matter differentiation to appear less distinct. Two cadavers with preserved grey‐white matter received acceptable scores for all image quality features for the protocol with a CTDIvol of 57.3 mGy, allowing organ dose savings ranging from 34% to 45%. One cadaver allowed for greater dose reduction for the protocol with a CTDIvol of 42 mGy.Conclusion:Efforts to optimize scan protocol should consider both dose and clinical image quality. This is made possible with postmortem subjects, whose brains are similar to patients, allowing for an investigation of ideal scan parameters. Radiologists at our institution accepted scan protocols acquired with lower scan parameters, with CTDIvol values closer to the American College of Radiology's (ACR) Achievable Dose level of 57 mGy.
PURPOSE Patient table height plays an important role in estimating patient skin dose for interventional radiology (IR) procedures, because the patient's skin location is dependent on the height of table. Variation in table height can lead to as much as 150% difference in skin dose for patient exams with similar air kerma meter readings. In our facility, IR procedural workflow was recently changed to require the IR physicians to confirm the patient table height before the procedure. The patient table height data was collected before and after this workflow change to validate the implementation of this practice. METHODS Table height information was analyzed for all procedures performed in three IR rooms, which were impacted by the workflow change, covering three months before and after the change (Aug 2015 to Jan 2016). In total, 442, 425, and 390 procedures were performed in these three rooms over this time period. There were no personnel or procedure assignment changes during the six-month period of time. Statistical analysis was performed for the average table height changes before and after the workflow change. RESULTS For the three IR rooms investigated, after the workflow change, the average table heights were increased by 1.43 cm (p=0.004084), 0.66 cm (p=0.187089), and 1.59 cm (p=0.002193), providing a corresponding estimated skin dose savings of 6.76%, 2.94% and 7.62%, respectively. After the workflow change, the average table height was increased by 0.95 cm, 0.63 cm, 0.55 cm, 1.07 cm, 1.12 cm, and 3.36 cm for the six physicians who routinely work in these three rooms. CONCLUSION Consistent improvement in table height settings has been observed for all IR rooms and all physicians following a simple workflow change. This change has led to significant patient dose savings by making physicians aware of the pre-procedure table position.
Purpose:This retrospective study analyzes the exposure history of emergency department (ED) patients undergoing head and cervical spine trauma computed tomography (CT) studies. This study investigated dose levels received by trauma patients and addressed any potential concerns regarding radiation dose issues.Methods:Under proper IRB approval, a cohort of 300 trauma cases of head and cervical spine trauma CT scans received in the ED was studied. The radiological image viewing software of the hospital was used to view patient images and image data. The following parameters were extracted: the imaging history of patients, the reported dose metrics from the scanner including the volumetric CT Dose Index (CTDIvol) and Dose Length Product (DLP). A postmortem subject was scanned using the same scan techniques utilized in a standard clinical head and cervical spine trauma CT protocol with 120 kVp and 280 mAs. The CTDIvol was recorded for the subject and the organ doses were measured using optically stimulated luminescent (OSL) dosimeters. Typical organ doses to the brain, thyroid, lens, salivary glands, and skin, based on the cadaver studies, were then calculated and reported for the cohort.Results:The CTDIvol reported by the CT scanner was 25.5 mGy for the postmortem subject. The average CTDIvol from the patient cohort was 34.1 mGy. From these metrics, typical average organ doses in mGy were found to be: Brain (44.57), Thyroid (33.40), Lens (82.45), Salivary Glands (61.29), Skin (47.50). The imaging history of the cohort showed that on average trauma patients received 26.1 scans over a lifetime.Conclusion:The average number of scans received on average by trauma ED patients shows that radiation doses in trauma patients may be a concern. Available dose tracking software would be helpful to track doses in trauma ED patients, highlighting the importance of minimizing unnecessary scans and keeping doses ALARA.
Purpose: New radiation dose reduction technologies are emerging constantly in the medical imaging field. The latest of these technologies, iterative reconstruction (IR) in CT, presents the ability to reduce dose significantly and hence provides great opportunity for CT protocol optimization. However, without effective analysis of image quality, the reduction in radiation exposure becomes irrelevant. This work explores the use of postmortem subjects as an image quality assessment medium for protocol optimizations in abdominal CT. Methods: Three female postmortem subjects were scanned using the Abdomen-Pelvis (AP) protocol at reduced minimum tube current and target noise index (SD) settings of 12.5, 17.5, 20.0, and 25.0. Images were reconstructed using two strengths of iterative reconstruction. Radiologists and radiology residents from several subspecialties were asked to evaluate 8 AP image sets including the current facility default scan protocol and 7 scans with the parameters varied as listed above. Images were viewed in the soft tissue window and scored on a 3-point scale as acceptable, borderline acceptable, and unacceptable for diagnosis. The facility default AP scan was identified to the reviewer while the 7 remaining AP scans were randomized and de-identified of acquisition and reconstruction details. The observers were also asked to comment on the subjective image quality criteria they used for scoring images. This included visibility of specific anatomical structures and tissue textures. Results: Radiologists scored images as acceptable or borderline acceptable for target noise index settings of up to 20. Due to the postmortem subjects’ close representation of living human anatomy, readers were able to evaluate images as they would those of actual patients. Conclusion: Postmortem subjects have already been proven useful for direct CT organ dose measurements. This work illustrates the validity of their use for the crucial evaluation of image quality during CT protocol optimization, especially when investigating the effects of new technologies.