Reject rate analysis is considered an integral part of a diagnostic radiography quality control (QC) program. A rejected image is a patient radiograph that was not presented to a radiologist for diagnosis and that contributes unnecessary radiation dose to the patient. Reject rates that are either too high or too low may suggest systemic department shortcomings in QC mechanisms. Due to the lack of standardization, reject data often cannot be easily compared between radiography systems from different vendors. The purpose of this report is to provide guidance to help standardize data elements that are required for comprehensive reject analysis and to propose data reporting and workflows to enable an effective and comprehensive reject rate monitoring program. Essential data elements, a proposed schema for classifying reject reasons, and workflow implementation options are recommended in this task group report.
Beginning with the advent of digital radiography systems in 1981, manufacturers of these systems provided indicators of detector exposure. These indicators were manufacturer-specific, and users in facilities with equipment from multiple manufacturers found it a challenge to monitor and manage variations in indicated exposure in routine clinical use. In 2008, a common definition of exposure index (EI) was realized in International Electrotechnical Commission (IEC) International Standard 62494-1 Ed. 1, which also introduced and defined the deviation index (DI), a number quantifying the difference between the detector EI for a given radiograph and the target exposure index (EIT ). An exposure index that differed by a constant from that established by the IEC and the concept of the deviation index also appear in American Association of Physicists in Medicine (AAPM) Report No. 116 published in 2009. The AAPM Report No. 116 went beyond the IEC standard in supplying a table (Table II in the report of TG-116) titled "Exposure Indicator DI Control Limits for Clinical Images," which listed suggested DI ranges and actions to be considered for each range. As the IEC EI was implemented and clinical DI data were gathered, concerns were voiced that the DI control limits published in the report of TG-116 were too strict and did not accurately reflect clinical practice. The charge of task group 232 (TG-232) and the objective of this final report was to investigate the current state of the practice for CR/DR Exposure and Deviation Indices based on AAPM TG 116 and IEC-62494, for the purpose of establishing achievable goals (reference levels) and action levels in digital radiography. Data corresponding to EI and DI were collected from a range of practice settings for a number of body parts and views (adults and pediatric radiographs) and analyzed in aggregate and separately. A subset of radiographs was also evaluated by radiologists based on criteria adapted from the European Guidelines on Quality Criteria for Diagnostic Radiographic Images from the European Commission. Analysis revealed that typical DI distribution was characterized by a standard deviation (SD) of 1.3-3.6 with mean DI values substantially different from 0.0, and less than 50% of DI values fell within the significant action limits proposed by AAPM TG-116 (-1.0 ≤ DI ≤ 1.0). Recommendations stemming from this analysis include targeting a mean DI value of 0.0 and action limits at ±1 and ±2 SD of the DI based on actual DI data of an individual site. EIT values, DI values, and associated action limits should be reviewed on an ongoing basis and optimization of DI values should be a process of continuous quality improvement with a goal of reducing practice variation.
PURPOSE To characterize the distribution of the deviation index (DI) in digital radiography practices across the United States. METHODS DI data was obtained from 10 collaborating institutions in the United States between 2012 and 2015. Each institution complied with the requirements of the Institutional Review Board at their site. DI data from radiographs of the body parts chest, abdomen, pelvis and extremity were analyzed for anteroposterior, posteroanterior, lateral, and decubitus views. The DI data was analyzed both in aggregate and stratified by exposure control method, image receptor technology, patient age, and participating site for each body part and view. The number of exposures with DI falling within previously published control limits for DI and descriptive statistics were calculated. RESULTS DI data from 505,930 radiographic exposures was analyzed. The number of exposures with DI falling within published control limits for DI varied from 10 to 20% for adult patients and 10 to 23% for pediatric patients for different body parts and views. Mean DI values averaged over other parameters for radiographs of the abdomen, chest, pelvis, and extremities ranged from 0.3 to 1.0, -0.6 to 0.5, 0.8, and -0.9 to 0.5 for the different adult views and ranged from -1.6 to -0.1, -0.3 to 0.5, -0.1, -0.2 to 1.4 for the different pediatric views, respectively (DI data was solicited only for anteroposterior view of pelvis). Standard deviation values of DI from individual sites ranged from 1.3 to 3.6 and 1.3 to 3.0 for the different adult and pediatric views, respectively. Also of interest was that target exposure indicators varied by up to a factor of 6 between sites for certain body parts and views. CONCLUSION Previously published DI control limits do not reflect the state of clinical practice in digital radiography. Mean DI and target exposure indicators are targets for quality improvement efforts in radiography.