Monte Carlo simulations were performed to extend existing neutron personal dose equivalent fluence-to-dose conversion coefficients to an energy of 250 MeV. Presently, conversion coefficients, H(p,slab)(10,alpha)/Phi, are given by ICRP-74 and ICRU-57 for a range of angles of radiation incidence (alpha = 0, 15, 30, 45, 60 and 75 degrees ) in the energy range from thermal to 20 MeV. Standard practice has been to base operational dose quantity calculations <20 MeV on the kerma approximation, which assumes that charged particle secondaries are locally deposited, or at least that charged particle equilibrium exists within the tally cell volume. However, with increasing neutron energy the kerma approximation may no longer be valid for some energetic secondaries such as protons. The Los Alamos Monte Carlo radiation transport code MCNPX was used for all absorbed dose calculations. Transport models and collision-based energy deposition tallies were used for neutron energies >20 MeV. Both light and heavy ions (HIs) (carbon, nitrogen and oxygen recoil nuclei) were transported down to a lower energy limit (1 keV for light ions and 5 MeV for HIs). Track energy below the limit was assumed to be locally deposited. For neutron tracks <20 MeV, kerma factors were used to obtain absorbed dose. Results are presented for a discrete set of angles of incidence on an ICRU tissue slab phantom.
Page Preface iii1 Radiation Measurements 111 Introduction 112 Materials and Equipment 1121 The Rando Phantom 1122 Radioactive Sources 1123 Philips SKYLight Gamma Camera 213 Radiation Measurements 42 Monte Carlo Simulations of Experimental Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 521 Methodology 5211 MCNP Model of Rando Phantom 52.1.2 Model of Philips SKYLight Gamma Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.2 Comparison of MCNP Simulations with Experimental Measurements . . . . . . . . . . . . . . 7References 8
CHELSI is a CsI-based portable spectrometer being developed at Los Alamos National Laboratory for use in high-energy neutron fields. Based on the inherent pulse shape discrimination properties of CsI(Tl), the instrument flags charged particle events produced via neutron-induced spallation events. Scintillation events are processed in real time using digital signal processing and a conservative estimate of neutron dose rate is made based on the charged particle energy distribution. A more accurate dose estimate can be made by unfolding the 2D charged particle versus pulse height distribution to reveal the incident neutron spectrum from which dose is readily obtained. A prototype probe has been assembled and data collected in quasi-monoenergetic fields at The Svedberg Laboratory (TSL) in Uppsala as well as at the Los Alamos Neutron Science Center (LANSCE). Preliminary efforts at deconvoluting the shape/energy data using empirical response functions derived from time-of-flight measurements are described.
The Health Physics Measurements Group at the Los Alamos National Laboratory (LANL) has initiated a study of neutron reference fields at selected US Department of Energy (DOE) calibration facilities. To date, field characterisation has been completed at five facilities. These fields are traceable to the National Institute for Standards and Technology (NIST) through either a primary calibration of the source emission rate or through the use of a secondary standard. However, neutron spectral variation is caused by factors such as room return, scatter from positioning tables and fixtures, source anisotropy and spectral degradation due to source rabbits and guide tubes. Perturbations from the ideal isotropic point source field may impact the accuracy of instrument calibrations. In particular, the thermal neutron component of the spectrum, while contributing only a small fraction of the conventionally true dose, can contribute a significant fraction of a dosemeter's response with the result that the calibration becomes facility-specific. A protocol has been developed to characterise neutron fields that relies primarily on spectral measurements with the Bubble Technology Industries (BTI) rotating neutron spectrometer (ROSPEC) and the LANL Bonner sphere spectrometer. The ROSPEC measurements were supplemented at several sites by the BTI Simple Scintillation Spectrometer probe, which is designed to extend the ROSPEC upper energy range from 5 to 15 MeV In addition, measurements were performed with several rem meters and neutron dosemeters. Detailed simulations were performed using the LANL MCNPX Monte Carlo code to calculate the magnitude of source anisotropy and scatter factors.
The intrinsic pulse shape discrimination properties of CsI(Tl) form the basis of a high-energy neutron (> 20MeV) spectrometer (CHELSI) currently being developed at LANL that shows promise in satisfying the requirements of an ideal survey meter; lightweight, portable and real time display of dose.Charged particle spallation products generated in the scintillator via neutron interactions are identified on the basis of pulse shape using digital pulse processing. Conservative estimates of dose rate can be given in real time based on count rates and pulse height distributions. More accurate dose measurements can be done off-line using unfolding methods to analyze stored pulse shape versus energy data.As a precursor to the development of a portable instrument, data has been obtained using a 1" x 1" CsI-based probe and a digital spectrometer. This system has been used to collect data on the 90 m flight path at the LANSCE/WNR facility at an average neutron energy of 335 MeV. The spectrometer has the capability, in addition to storing individual waveforms for later analysis, of recording time-of-flight data and calculating a pulse shape parameter and pulse height for each scintillation event in real time. Combining these data with traditional multichannel analyzer data has yielded a set of empirical response functions with respect to neutron energy. Analysis of the charged particle spectra has yielded an overall average count rate of 0.12 +/- 0.02 cps/mu Sv h for a 1" x 1" CsI(Tl) scintillator in this neutron field. (c) 2006 Elsevier B.V. All rights reserved.
Radiation sources producing short pulses of photon radiation are widespread. Such sources include electron linear accelerators and field emission impulse generators. It is often desirable to measure leakage and skyshine radiation for these sources in real time and at environmental levels as low as 0.02 microSv per pulse. This note provides an overview of the design and performance of a commercial, real-time, pulsed photon dosemeter (PPD) capable of single-pulse dose measurements over the range from 0.02 to 20 microSv. The PPD may also be operated in a multiple-pulse mode that integrates the dose from a train of pulses over a 3 s period. A pulse repetition rate of up to 300 Hz is accommodated.
A simple dosimeter made of a sulfur tablet, bare and cadmium covered indium foils and a cadmium covered copper foil has been modeled using MCNP5. Studies of the model without phantom or other confounding factors have shown that the cross sections and fluence-to-dose factors generated by the Monte Carlo method agree with those generated by analytic expressions for the high energy component. The threshold cross sections for the detectors on a phantom were calculated. The resulting doses assigned agree well with exposures made to three critical assemblies. In this study the angular dependence on a phantom is studied and compared with measurements taken on the GODIVA reactor. The dosimeter positions on the phantom are facing the source, on the back and the side. In previous papers the modeling of a simple dosimeter made of a sulfur tablet, bare and cadmium covered indium foils and a cadmium covered copper foil has been modeled using MCNP5. The conclusion made was that most of the neutron dose from criticality assemblies results from the high energy neutron fluences determined by the sulfur and indium detectors. The results using doses measured from the GODIVA, SHEBA, and bare and lead shielded SILENE reactors confirmed this. Themore » angular dependence of an accident dosemeter is of interest in evaluating the exposure of personnel. To investigate this effect accident dosemeters were placed on a phantom and exposed to the GODIVA reactor at phantom orientations of 0{sup o}, 45{sup o}, 90{sup o}, 135{sup o}, and 180{sup o} to the assembly center line.« less
Conventional neutron rem meters currently in use are based on 1960's technology that relies on a large neutron moderator assembly surrounding a thermal detector to achieve a rem-like response function over a limited energy range. Such rem meters present an ergonomic challenge, being heavy and bulky, and have caused injuries during radiation protection surveys. Another defect of traditional rem meters is a poor high-energy response above 10 MeV, which makes them unsuitable for applications at high-energy accelerator facilities. Proton Recoil Scintillator-Los Alamos (PRESCILA) was developed as a low-weight (2 kg) alternative capable of extended energy response, high sensitivity, and moderate gamma rejection. An array of ZnS(Ag) based scintillators is located inside and around a Lucite light guide, which couples the scintillation light to a sideview bialkali photomultiplier tube. The use of both fast and thermal scintillators allows the energy response function to be optimized for a wide range of operational spectra. The light guide and the borated polyethylene frame provide moderation for the thermal scintillator element. The scintillators represent greatly improved versions of the Hornyak and Stedman designs from the 1950's, and were developed in collaboration with Eljen Technology. The inherent pulse height advantage of proton recoils over electron tracks in the phosphor grains eliminates the need for pulse shape discrimination and makes it possible to use the PRESCILA probe with standard pulse height discrimination provided by off-the-shelf health physics counters. PRESCILA prototype probes have been extensively tested at both Los Alamos and the German Bureau of Standards, Physikalisch-Technische Bundesanstalt. Test results are presented for energy response, directional dependence, linearity, sensitivity, and gamma rejection. Initial field tests have been conducted at Los Alamos and these results are also given. It is concluded that PRESCILA offers a viable, ergonomically superior, alternative to traditional rem meters that is effective for a wide range of neutron fields. The probe is capable of excellent sensitivity (40 counts per minute per microSv h-1 for 241AmBe) and extended energy response to beyond 20 MeV. Directional response is uniform (+/-15%) over a wide range of energies. Response linearity has been characterized to over 20 mSv h-1. Gamma rejection is effective in gamma fields up to 2 mSv h-1. The PRESCILA technology has been commercialized and is now offered under license by Ludlum Measurements, Inc.
Initial calibration of a multisphere spectroscopy system has been completed at Los Alamos National Laboratory using four standard calibration scenarios. Spectrum unfolding was performed using three methods of constructing the default spectrum: simple parameter models, Monte Carlo calculations and physical measurement. Comparisons of the resulting spectra for each solution method are presented. Implications of the spectral solutions upon dosemeter characterisation are addressed.
The design of a portable high-energy (20-800 MeV) neutron spectrometer based on CsI or BaF2 is described. The particle discrimination properties of these scintillators allow the light-ion spallation products (p, d, t and alpha) from neutron interactions to be identified uniquely. One or more of the resulting pulse-height spectra can be unfolded to reveal the incident neutron spectrum. Dosimetric quantities can then be calculated based on the unfolded spectrum. Due to the high stopping power of these scintillators, modest-sized crystals are suitable for this application. Combined with advances in electronics, a lightweight instrument capable of on-line particle discrimination with a real-time display of neutron-induced count rate is feasible. Preliminary experimental data are presented, and the importance of validating MCNPX-generated response functions is discussed. A brief discussion on future work follows.
Several neutron spectrometers manufactured by Bubble Technology Industries (BTI) were tested and evaluated in a variety of neutron fields. Findings and conclusions are presented for the following BTI instruments: a modification of the Rotational Spectrometer (ROSPEC) that includes a thermal and epithermal capability, the Simple Scintillation Spectrometer that is used in conjunction with the ROSPEC to extend its high-energy range, and the MICROSPEC N-Probe which is capable of providing a crude spectrum over the energy range from thermal to 18MeV. The main objective of these measurements was to determine the accuracy of both the energy spectrum and dose equivalent information generated by these devices. In addition, the dose response of the Wide-Energy Neutron Detection Instrument (WENDI-II) was measured in all neutron fields relative to a bare 252Cf calibration. The performance of the WENDI-II rem meter was compared to the dose information generated by the neutron spectrometers. The instruments were irradiated to bare 252Cf and 241AmBe sources, and in a series of moderated 252Cf fields using a standard D2O sphere and a set of polyethylene spheres. The measured spectra were benchmarked with a set of detailed Monte Carlo calculations with the same energy bin structure as that of the instruments under test. These calculations allowed an absolute comparison to be made with the measurements on a bin by bin basis. The simulations included the effects of room return and source anisotropy.
Neutron rem meters are routinely used for real-time field measurements of neutron dose equivalent where neutron spectra are unknown or poorly characterized. These meters are designed so that their response per unit fluence approximates an appropriate fluence-to-dose conversion function. Typically, a polyethylene moderator assembly surrounds a thermal neutron detector, such as a BF3 counter tube. Internal absorbers may also be used to further fine-tune the detector response to the shape of the desired fluence conversion function. Historical designs suffer from a number of limitations. Accuracy for some designs is poor at intermediate energies (50 keV-250 keV) critical for nuclear power plant dosimetry. The well-known Andersson-Braun design suffers from angular dependence because of its lack of spherical symmetry. Furthermore, all models using a pure polyethylene moderator have no useful high-energy response, which makes them inaccurate around high-energy accelerator facilities. This paper describes two new neutron rem meter designs with improved accuracy over the energy range from thermal to 5 GeV. The Wide Energy Neutron Detection Instrument (WENDI) makes use of both neutron generation and absorption to contour the detector response function. Tungsten or tungsten carbide (WC) powder is added to a polyethylene moderator with the expressed purpose of generating spallation neutrons in tungsten nuclei and thus enhance the high-energy response of the meter beyond 8 MeV. Tungsten's absorption resonance structure below several keV was also found to be useful in contouring the meter's response function. The WENDI rem meters were designed and optimized using the Los Alamos Monte Carlo codes MCNP, MCNPX, and LAHET. A first generation prototype (WENDI-I) was built in 1995 and its testing was completed in 1996. This design placed a BF3 counter in the center of a spherical moderator assembly, whose outer shell consisted of 30% by weight WC in a matrix of polyethylene. A borated silicone rubber (5% boron by weight) absorber covered an inner polyethylene sphere to control the meter's response at intermediate energies. A second generation design (WENDI-II) was finalized and tested in 1999. It further extended the high-energy response beyond 20 MeV, increased sensitivity, and greatly facilitated the manufacturing process. A 3He counter tube is located in the center of a cylindrical polyethylene moderator assembly. Tungsten powder surrounds the counter tube at an inner radius of 4 cm and performs the double duty of neutron generation above 8 MeV and absorption below several keV. WENDI-II is suitable for field use as a portable rem meter in a variety of work place environments, and has been recently commercialized under license by Eberline Instruments, Inc. and Ludlum Measurements, Inc. Sensitivity is about a factor of 12 higher than that of the Hankins Modified Sphere (Eberline NRD meter) in a bare 252Cf field. Additionally, the energy response for WENDI-II closely follows the contour of the Ambient Dose Equivalent per unit fluence function [H'(10)/phi] above 0.1 MeV. Its energy response at 500 MeV is approximately 15 times higher than that of the Hankins and Andersson-Braun meters. Measurements of the energy and directional response of the improved meter are presented and the measured response function is shown to agree closely with the predictions of the Monte Carlo simulations in the range from 0.144 MeV to 19 MeV.
The Plutonium Facility at Los Alamos uses Eberline RMS-II monitors as part of its criticality accident alarm system. The monitors are designed to accommodate a wide spectrum of criticality accident scenarios from both liquid and metal supercriticalities. A series of experiments were performed at the Los Alamos Critical Experiments Facility to evaluate the response of the monitor to a wide range of radiation transients. The primary objectives of the evaluation were to verify reliable and prompt latching of the criticality alarms in accordance with the performance criteria of ANSI/ANS-8.3-1986. The RMS-II monitor performed reliably during all of the experiments and was found to meet specific ANSI/ANS performance criteria for latching.