Abstract Background High ischemic mitral regurgitation (IMR) recurrence rates continue to plague IMR repair with undersized ring annuloplasty. We have previously shown that pre-repair three-dimensional echocardiography (3DE) analysis is highly predictive of IMR recurrence. The objective of this study was to determine the quantitative change in 3DE annular and leaflet tethering parameters immediately after repair and to determine if intraoperative post-repair 3DE parameters would be able to predict IMR recurrence 6 months after repair. Methods Intraoperative pre- and post-repair transesophageal real-time 3DE was performed in 35 patients undergoing undersized ring annuloplasty for IMR. An advanced modeling algorhythm was used to assess 3D annular geometry and regional leaflet tethering. IMR recurrence (≥ grade 2) was assessed with transthoracic echocardiography 6 months after repair. Results Annuloplasty significantly reduced septolateral diameter, commissural width, annular area, and tethering volume and significantly increased all segmental tethering angles (except A2). Intraoperative post-repair annular geometry and leaflet tethering did not differ significantly between patients with recurrent IMR (n = 9) and patients with non-recurrent IMR (n = 26). No intraoperative post-repair predictors of IMR recurrence could be identified. Conclusions Undersized ring annuloplasty changes mitral geometry acutely, exacerbates leaflet tethering, and generally fixes IMR acutely, but it does not always fix the delicate underlying chronic problem of continued left ventricular dilatation and remodeling. This may explain why pre-repair 3D valve geometry (which reflects chronic left ventricular remodeling) is highly predictive of recurrent IMR, whereas immediate post-repair 3D valve geometry (which does not completely reflect chronic left ventricular remodeling anymore) is not.
Objective Durability of mitral valve (MV) repair for functional mitral regurgitation (FMR) remains suboptimal. We sought to create a highly reproducible, quantitative ex vivo model of FMR that functions as a platform to test novel repair techniques. Methods Fresh swine hearts ( n = 10) were pressurized with air to a left ventricular pressure of 120 mmHg. The left atrium was excised and the altered geometry of FMR was created by radially dilating the annulus and displacing the papillary muscle tips apically and radially in a calibrated fashion. This was continued in a graduated fashion until coaptation was exhausted. Imaging of the MV was performed with a 3-dimensional (3D) structured-light scanner, which records 3D structure, texture, and color. The model was validated using transesophageal echocardiography in patients with normal MVs and severe FMR. Results Compared to controls, the anteroposterior diameter in the FMR state increased 32% and the annular area increased 35% ( P < 0.001). While the anterior annular circumference remained fixed, the posterior circumference increased by 20% ( P = 0.026). The annulus became more planar and the tenting height increased 56% (9 to 14 mm, P < 0.001). The median coaptation depth significantly decreased (anterior leaflet: 5 vs 2 mm; posterior leaflet: 7 vs 3 mm, P < 0.001). The ex vivo normal and FMR models had similar characteristics as clinical controls and patients with severe FMR. Conclusions This novel quantitative ex vivo model provides a simple, reproducible, and inexpensive benchtop representation of FMR that mimics the systolic valvular changes of patients with FMR.
Although studied for many years, there remain continued gaps in our fundamental understanding of cardiac kinematics, such as the nature and extent of heart wall volumetric changes that occur over the cardiac cycle. Such knowledge is especially important for accurate in silico simulations of cardiac pathologies and in the development of novel therapies for their treatment. A prime example is myocardial infarction (MI), which induces profound, regionally variant maladaptive remodeling of the left ventricle (LV) wall. To address this problem, we conducted an in vivo fiduciary marker-based study in an established ovine model of MI to generate detailed, time-evolving transmural in vivo volumetric measurements of LV free wall deformations in the normal state, as well as up to 12 h post-MI. This was accomplished using a transmural array of sonomicrometry crystals that acquired fiducial positions at ∼250 Hz with a positional accuracy of ∼0.1 mm, covering the entire infarct, border, and remote zones. A convex-hull method was used to directly calculate the Jacobian J(t)=Δv(t)/ΔVED from sonocrystal positions over the entire cardiac cycle, where ΔV is the volume of each convex polyhedral at end diastole (ED) (typically ∼1 cc). We demonstrated significant in vivo compressibility in normal functioning LV free wall myocardium, with JES=0.85±0.07 at end systole (ES). We also observed substantial regional variations, with the largest reduction in local myocardial tissue volume during systole in the base region accompanied by substantial transmural gradients. These patterns changed profoundly following loss of perfusion post-MI, with the apical region showing the greatest loss of volume reduction at ES. To verify that the sonocrystals did not affect local volumetric measurements, JES measures were also verified by non-invasive magnetic resonance imaging, exhibiting very similar changes in regional volume. We note that while our estimates of regional compressibility were in close agreement with the values previously reported for large animals, ranging from 5% to 20%, the direct, comprehensive measurements of wall compressibility presented herein improved on the limitations of previous reports. These limitations included dependency on the small local volumes used for analysis and often indirect measurement of compressibility. Our novel findings suggest that proper accounting for the myocardial effective compressibility at the ∼1 cc volume scale can improve the accuracy of existing kinematic indices, such as wall thickening and axial shortening, and simulations of LV remodeling following MI.
Objectives: Repair for ischemic mitral regurgitation with undersized annuloplasty is characterized by high recurrence rates. We sought to determine the value of pre-repair 3-dimensional echocardiography over 2-dimensional echocardiography in predicting recurrence at 6 months.Methods: Intraoperative transesophageal 2-dimensional echocardiography and 3-dimensional echocardiography were performed in 50 patients undergoing undersized annuloplasty for ischemic mitral regurgitation. Two-dimensional echocardiography annular diameter and tethering parameters were measured in the apical 2-and 4-chamber views. A customized protocol was used to assess 3-dimensional annular geometry and regional leaflet tethering. Recurrence (grade >= 2) was assessed with 2-dimensional transthoracic echocardiography at 6 months.Results: Preoperative 2- and 3-dimensional annular geometry were similar in all patients with ischemic mitral regurgitation. Preoperative 2- and 3-dimensional leaflet tethering were significantly higher in patients with recurrence (n = 13) when compared with patients without recurrence (n = 37). Multivariate logistic regression revealed preoperative 2-dimensional echocardiography posterior tethering angle as an independent predictor of recurrence with an optimal cutoff value of 32.0 degrees (area under the curve, 0.81; 95% confidence interval, 0.68-0.95; P = .002) and preoperative 3-dimensional echocardiography P3 tethering angle as an independent predictor of recurrence with an optimal cutoff value of 29.9 degrees (area under the curve, 0.92; 95% confidence interval, 0.84-1.00; P < .001). The predictive value of the 3-dimensional geometric multivariate model can be augmented by adding basal aneurysm/dyskinesis (area under the curve, 0.94; 95% confidence interval, 0.87-1.00; P < .001).Conclusions: Preoperative 3-dimensional echocardiography P3 tethering angle is a stronger predictor of ischemic mitral regurgitation recurrence after annuloplasty than preoperative 2-dimensional echocardiography posterior tethering angle, which is highly influenced by viewing plane. In patients with a preoperative P3 tethering angle of 29.9 degrees or larger (especially when combined with basal aneurysm/dyskinesis), chordal-sparing valve replacement should be strongly considered.
Background: Mitral valve repair is the preferred treatment for ischemic mitral regurgitation (IMR); however, the high incidence of early failure emphasizes the need for parameters to identify patients at high risk for recurrent IMR. Using real-time (rt-3DE), regional mitral valve parameters were compared to better characterize recurrent and non-recurrent IMR valves. Methods: Pre-operative three-dimensional echocardiography (3DE) images of 48 IMR valves (35 non-recurrent and 13 recurrent) and 18 normal valves were acquired. Six month follow-up echoes of the IMR patients were obtained to assess competency of the valve post repair. Valve images were segmented as previously published. Customized algorithms were used to determine the septolateral diameter (SLD), mitral annular area (MAA), mitral valve tenting volume (MVTv) referring to the volume enclosed by the annular plane and mitral leaflets, mitral valve tenting index (MVTI) calculated by the MVTv divided by MAA, and tethering angles for each leaflet regi...
Background. Using human mitral valve (MV) models derived from three-dimensional echocardiography, finite element analysis was used to predict mechanical leaflet and chordal stress. Subsequently, valve geometries were altered to examine the effects on stresses of the following: (1) varying coaptation area; (2) varying noncoapted leaflet tissue area; and (3) varying interleaflet coefficient of friction (11).Methods. Three human MV models were loaded with a transvalvular pressure of 80 mm Hg using finite element analysis. Initially leaflet coaptation was set to 10%, 50%, or 100% of actual coaptation length to test the influence of coaptation length on stress distribution. Next, leaflet surface areas were augmented by 1% overall and by 2% in the noncoapted "belly" region to test the influence of increased leaflet billowing without changing the gross geometry of the MV. Finally, the coefficient of friction between the coapted leaflets was set to mu = 0, 0.05, or 0.3, to assess the influence of friction on MV function.Results. Leaflet coaptation length did not affect stress distribution in either the coapted or noncoapted leaflet regions; peak leaflet stress was 0.36 +/- 0.17 MPa at 100%, 0.35 +/- 0.14 MPa at 50%, and 0.35 +/- 0.15 MPa at 10% coaptation lengths (p = 0.85). Similarly, coaptation length did not affect peak chordal tension (p = 0.74). Increasing the noncoapted leaflet area decreased the peak valvular stresses by 5 +/- 2% (p = 0.02). Varying the coefficient of friction between leaflets did not alter leaflet or chordal stress distribution (p = 0.18).Conclusions. Redundant MV leaflet tissue reduces mechanical stress on the noncoapted leaflets; the extent of coaptation or frictional interleaflet interaction does not independently influence leaflet stresses. Repair techniques that increase or preserve noncoapted leaflet area may decrease mechanical stresses and thereby enhance repair durability. (Ann Thorac Surg 2012;93:1141-5) (C) 2012 by The Society of Thoracic Surgeons
BACKGROUND:Restoring leaflet coaptation is the primary objective in repair of ischemic mitral regurgitation (IMR). The common practice of placing an undersized annuloplasty ring partially achieves this goal by correcting annular dilation; however, annular reduction has been demonstrated to exacerbate posterior leaflet tethering. Using a sheep model of IMR, we tested the hypothesis that posterior leaflet augmentation (PLA) combined with standard annuloplasty sizing increases leaflet coaptation more effectively than undersized annuloplasty alone.METHODS:Eight weeks after posterobasal myocardial infarction, 15 sheep with 2+ or greater IMR underwent annuloplasty with either a 24-mm annuloplasty ring (24-mm group, n = 5), 30-mm ring (30-mm group, n = 5), or 30-mm ring with concomitant augmentation of the posterior leaflet (PLA group, n = 5). Using three-dimensional echocardiography, postrepair coaptation zone and posterior leaflet mobility were assessed.RESULTS:Leaflet coaptation length after repair was greater in the PLA group (4.1 ± 0.3 mm) and the 24-mm group (3.8 ± 0.5 mm) as compared with the 30-mm group (2.7 ± 0.6 mm, p < 0.01). Leaflet coaptation area was significantly greater in the PLA group (121.5 ± 6.6 mm(2)) as compared with the 30-mm group (77.5 ± 17.0 mm(2)) or the 24-mm group (92.5 ± 17.9 mm(2), p < 0.01). Posterior leaflet mobility was significantly greater in the PLA group as compared with the 30-mm group or the 24-mm group.CONCLUSIONS:Posterior leaflet augmentation combined with standard-sized annuloplasty enhances leaflet coaptation more effectively than either standard-sized annuloplasty or undersized annuloplasty alone. Increased leaflet coaptation after PLA provides redundancy to IMR repair, and may decrease incidence of both recurrent IMR and mitral stenosis.
Background A comprehensive three-dimensional echocardiography based approach is applied to preoperative mitral valve (MV) analysis in patients with ischemic mitral regurgitation (IMR). This method is used to characterize the heterogeneous nature of the pathologic anatomy associated with IMR. Methods Intraoperative real-time three-dimensional transesophageal echocardiograms of 18 patients with IMR (10 with anterior, 8 with inferior infarcts) and 17 patients with normal MV were analyzed. A customized image analysis protocol was used to assess global and regional determinants of annular size and shape, leaflet tethering and curvature, relative papillary muscle anatomy, and anatomic regurgitant orifice area. Results Both mitral annular area and MV tenting volume were increased in the IMR group as compared with patients with normal MV (mitral annular area=1,065 ± 59 mm2 versus 779 ± 44 mm2, p = 0.001; and MV tenting volume=3,413 ± 403 mm3 versus 1,696 ± 200 mm3, p = 0.001, respectively). Within the IMR group, patients with anterior infarct had larger annuli (1,168 ± 99 mm2) and greater tenting volumes (4,260 ± 779 mm3 versus 2,735 ± 245 mm3, p = 0.06) than the inferior infarct subgroup. Papillary-annular distance was increased in the IMR group relative to normal; these distances were largest in patients with anterior infarcts. Whereas patients with normal MV had very consistent anatomic determinants, annular shape and leaflet tenting distribution in the IMR group were exceedingly variable. Mean anatomic regurgitant orifice area was 25.8 ± 3.0 mm2, and the number of discrete regurgitant orifices varied from 1 to 4. Conclusions Application of custom analysis techniques to three-dimensional echocardiography images allows a quantitative and systematic analysis of the MV, and demonstrates the extreme variability in pathologic anatomy that occurs in patients with severe IMR.
BACKGROUND:Real-time three-dimensional (3D) echocardiography has the ability to construct quantitative models of the mitral valve (MV). Imaging and modeling algorithms rely on operator interpretation of raw images and may be subject to observer-dependent variability. We describe a comprehensive analysis technique to generate high-resolution 3D MV models and examine interoperator and intraoperator repeatability in humans. METHODS:Patients with normal MVs were imaged using intraoperative transesophageal real-time 3D echocardiography. The annulus and leaflets were manually segmented using a TomTec Echo-View workstation. The resultant annular and leaflet point cloud was used to generate fully quantitative 3D MV models using custom Matlab algorithms. Eight images were subjected to analysis by two independent observers. Two sequential images were acquired for 6 patients and analyzed by the same observer. Each pair of annular tracings was compared with respect to conventional variables and by calculating the mean absolute distance between paired renderings. To compare leaflets, MV models were aligned so as to minimize their sum of squares difference, and their mean absolute difference was measured. RESULTS:Mean absolute annular and leaflet distance was 2.4±0.8 and 0.6±0.2 mm for the interobserver and 1.5±0.6 and 0.5±0.2 mm for the intraobserver comparisons, respectively. There was less than 10% variation in annular variables between comparisons. CONCLUSIONS:These techniques generate high-resolution, quantitative 3D models of the MV and can be used consistently to image the human MV with very small interoperator and intraoperator variability. These data lay the framework for reliable and comprehensive noninvasive modeling of the normal and diseased MV.
Introduction: There is consensus that mitral valve repair procedures for ischemic mitral regurgitation (IMR) should employ undersized annuloplasty rings to restore normal annular area. Some surgeon...
Background Myxomatous disease is generally considered to be a disease of the leaflets, with preserved annular function. As such, flexible annuloplasty devices are routinely utilized to stabilize th...
Three-dimensional (3-D) echocardiography allows the generation of anatomically correct and time-resolved geometric mitral valve (MV) models. However, as imaged in vivo, the MV assumes its systolic geometric configuration only when loaded. Customarily, finite element analysis (FEA) is used to predict material stress and strain fields rendered by applying a load on an initially unloaded model. Therefore, this study endeavors to provide a framework for the application of in vivo MV geometry and FEA to MV physiology, pathophysiology, and surgical repair. We hypothesize that in vivo MV geometry can be reasonably used as a surrogate for the unloaded valve in computational (FEA) simulations, yielding reasonable and meaningful stress and strain magnitudes and distributions. Three experiments were undertaken to demonstrate that the MV leaflets are relatively nondeformed during systolic loading: 1) leaflet strain in vivo was measured using sonomicrometry in an ovine model, 2) hybrid models of normal human MVs as constructed using transesophageal real-time 3-D echocardiography (rt-3DE) were repeatedly loaded using FEA, and 3) serial rt-3DE images of normal human MVs were used to construct models at end diastole and end isovolumic contraction to detect any deformation during isovolumic contraction. The average linear strain associated with isovolumic contraction was 0.02 ± 0.01, measured in vivo with sonomicrometry. Repeated loading of the hybrid normal human MV demonstrated little change in stress or geometry: peak von Mises stress changed by <4% at all locations on the anterior and posterior leaflets. Finally, the in vivo human MV deformed minimally during isovolumic contraction, as measured by the mean absolute difference calculated over the surfaces of both leaflets between serial MV models: 0.53 ± 0.19 mm. FEA modeling of MV models derived from in vivo high-resolution truly 3-D imaging is reasonable and useful for stress prediction in MV pathologies and repairs.
Objective: Degenerative mitral regurgitation is common and frequently addressed using mitral valve repair. Using intraoperative real-time three-dimensional echocardiography (rt-3DE) and specially developed software, we describe a high resolution, quantitative geometric model of human mitral valves with flail mid-posterior (P2) leaflet segments. Methods: Intraoperative transesophageal echocardiography (TEE) using a Philips ie33 ultrasound module and x7-2t rt-3DE probe was performed prior to surgical repair of degenerative mitral regurgitation in 15 patients with P2 flail. Images were imported to an offline workstation for analysis with Tomtec Echo-View. Fully 3D models of the annuli and leaflets under loaded conditions (midsystole) were created using custom Matlab algorithms. Anatomic regurgitant orifice (ARO) was traced by planimetry in the 3D rendered model of the mitral valve. Results: 3D models demonstrated significant heterogeneity in leaflet shape and regurgitant orifice (figure). Three distinct morp...
OBJECTIVES:Using 3-dimensional echocardiography in conjunction with novel geometric modeling and rendering techniques, we have developed a high-resolution, quantitative, 3-dimensional methodology for imaging the human mitral valve. Leaflet and annular geometry are important determinants of mitral valve stress. Repair techniques that optimize valvular geometry will reduce stress and potentially increase repair durability. The development of such procedures will require image-processing methodologies that provide a quantitative description of 3-dimensional valvular geometry.METHODS:Ten healthy adult subjects underwent mitral valve imaging with real-time 3-dimensional echocardiography. By using specially designed image analysis software, multiple valvular geometric parameters, including 2- and 3-dimensional leaflet curvature, leaflet surface area, annular height, intercommissural width, septolateral annular diameter, and annular area were determined for each subject. Image-rendering techniques that allow for the clear and concise presentation of this detailed information are also presented.RESULTS:Although 3-dimensional annular and leaflet geometry were found to be highly conserved between healthy human subjects in general, substantial intrasubject and intersubject regional geometric heterogeneity was observed in the midposterior leaflet, the region most commonly involved in leaflet flail in subjects with myxomatous disease.CONCLUSIONS:The image-processing and graphic-rendering techniques that we have developed can be used to provide a complete description of 3-dimensional mitral valve geometry in human subjects. Widespread application of these techniques to healthy subjects and patients with mitral valve disease will provide insight into the geometric basis of both valvular pathology and repair durability.
Objectives: High leaflet and chordal stresses contribute to recurrent mitral regurgitation after repair procedures. We hypothesized that a saddle-shaped annuloplasty ring would reduce leaflet stress compared to a similarly sized flat annuloplasty ring. To test this hypothesis we used a novel 3D echocardiographically-based finite element modeling (FEM) technique for quantifying regional mitral valve stress. Methods: Real-time 3D echocardiography was performed in 8 sheep before and after placement of either a 30mm flat annuloplasty (n=4) or a 30mm saddle-shaped annuloplasty. Full-volume data sets of the MV were obtained using an IE33 platform(Philips Medical Systems, Andover, Massachusetts) and exported to Cardio-View (Tomtec Imaging Systems, Munich, Germany) for image analysis. Individual leaflet data were then interpolated using Matlab (The Mathworks, Natick, Massachusetts). Triangulated leaflet surfaces were extracted and the data imported into a commercial finite element program (ABAQUS/Explicit 6.3, HKS Inc. Pawtucket, RI) to quantify regional stress distributions in all segments (P1, P2, P3 and A1, A2, A3) of the MV. Results: Peak anterior and posterior leaflet stresses after flat annuloplasty placement were 0.20±0.001MPa and 21±0.02MPa. Peak anterior and posterior leaflet stress after saddle-shaped annuloplasty placement was 0.19±0.02MPa and 13±0.01MPa (p<0.05 for the posterior leaflet stresses) Conclusions: Saddle-shaped annuloplasty design results in greater stress reduction in the posterior leaflet than standard flat annuloplasty rings and may, therefore, result in more durable repairs. This research has received full or partial funding support from the American Heart Association, AHA Great Rivers Affiliate (Delaware, Kentucky, Ohio, Pennsylvania & West Virginia).
BACKGROUND:The effect of mitral leaflet curvature on stress reduction is an important mechanism in optimizing valve function. We hypothesize that annuloplasty ring shape could directly influence leaflet curvature and, potentially, repair durability. We describe an echocardiographically based methodology for quantifying mitral valve geometry and its application to the characterization of ovine mitral valve geometry before and after implantation of an annuloplasty ring. METHODS:Multiple mitral annular and leaflet geometric variables were calculated for 8 naïve adult male sheep using real-time three-dimensional echocardiographic images. These indexes were recalculated after annuloplasty using a 30-mm Carpentier-Edward Physio ring (n = 4; Edwards Lifesciences, Irvine, CA) or a 30-mm saddle ring (n = 4). RESULTS:After implantation of the Physio ring, the annular height to commissural width ratio (AHCWR) decreased from 19.4% +/- 2.3% to 11.1% +/- 2.5% (p = 0.06). After implantation of the saddle ring, AHCWR increased from 19.6% +/- 1.3% to 24.3% +/- 1.3% (p < 0.05). Statistically significant increases in three-dimensional Gaussian curvature occurred after implantation within six defined leaflet regions (A1 to A3, P1 to P3) of the saddle ring but only within the P1 and P3 leaflet regions with the Physio ring. CONCLUSIONS:Annuloplasty ring shape affects leaflet curvature. Implantation of a saddle ring reflecting normal human annular geometry augmented ovine annular nonplanarity and increased three-dimensional leaflet curvature across the entire mitral valve surface. The Physio ring decreased annular nonplanarity and increased leaflet curvature only across limited regions of the posterior leaflet. These findings confirm the hypothesis that ring design influences leaflet curvature.